17. Signal Id Oscillator Adjustment; Signal Id Oscillator Adjustment Setup - HP 8560A Service Manual

Spectrum analyzer
Table of Contents

Advertisement

17. Signal ID Oscillator Adjustment

Assembly Adjusted
Al5 RF Assembly
Related Performance Test
There is no related performance test for this adjustment procedure.
Description
The frequency range of the 298 MHz Signal ID Oscillator is determined by counting the
10.7 MHz IF as A15C629 is rotated through its range of adjustment. The Signal ID Oscillator
is then set to the frequency determined by the following equation:
Oscillator frequency = 298 MHz -
SIGNAL
ANALYZER
Figure 2-20. Signal ID Oscillator Adjustment Setup
Equipment
Microwave Frequency Counter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP 5343A
Spectrum Analyzer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP 8566A/B
Adapters
Type N (m) to BNC (f) (2 required) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1250-1476
BNC Tee (f,m,f) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ...1250-0781
FREQUENCY
COUNTER
ADAPTER
TEE
17. Signal ID Oscillator Adjustment
Oscillator frequency range
4
SPECTRUM
ANALYZER
Adjustment Procedures 2-63

Advertisement

Table of Contents
loading

Table of Contents