Recommended Test Equipment - HP 8560A Service Manual

Spectrum analyzer
Table of Contents

Advertisement

Instrument
Synthesized
Frequency Range: 10 MHz to 7 GHz
Sweeper
Frequency Accuracy (CW): 1 x 10eg/day
(two required)
Leveling Modes: Internal & External
Modulation Modes: AM & Pulse
Power Level Range: -80 to +16 dBm
Synthesizer/
Frequency Range: 200 Hz to 80 MHz
Level Generator
Frequency Accuracy: 1 x 10e7/month
Flatness: f0.15 dB
Attenuator Accuracy: <&0.09 dB
Synthesized
Frequency Range: 5 MHz to 2.5 GHz
Signal Generator
Residual SSB Phase Noise at 10 kHz offset
(320 MHz <fc <640 MHz): <-131 dBc/Hz
Pulse/Function
Frequency Range: 10 kHz to 50 MHz
Generator
Pulse Width: 200 ns;
Output Amplitude: 5 V peak-to-peak
Functions: Pulse & Triangle
TTL Sync Output
Frequency Range: 1 MHz to 200 MHz
AM/FM
Frequency Modulation Mode
Signal Generator
Modulation Oscillator Frequency: 1 kHz
FM Peak Deviation: 5 kHz
Microwave
Frequency Range: 9 MHz to 6.9 GHz
Frequency Counter Time base Accuracy (Aging): <5
External Frequency Reference Input
Universal
Modes: TI AFB, Frequency Count
Time Interval Measurement Range: 20 ms to 75 s
Counter
Frequency Measurement Range: 17 kHz to 11 MHz
Frequency Resolution: 1 MHz at 10 MHz
External Frequency Reference Input
Time base accuracy (Aging): <3
Bandwidth (3 dB): dc to 100 MHz
Oscilloscope
Minimum Vertical Deflection Factor: <2 mV/div
Frequency Range: 4 kHz to 7 GHz
Spectrum Analyzer
Typical Residual FM: <l Hz Pk-to-Pk in 100 ms
(Fundamental Mixing)
* Part of Microwave Workstation
P = Performance Tests; A = Adjustments; M = Test & Adjustment Module; T = Troubleshooting;
V = Operation Verification
Table 1-4. Recommended Test Equipment
Critical Specifications
for Equipment Substitution
Recommended
Model
HP 8340A*
HP 3335A*
HP 8663A
HP 8116A
HP 8640B
HP 5343A*
Option 001
x
HP
x
HP
HP 8566A/B
General Information
Use
P
A
P

Advertisement

Table of Contents
loading

Table of Contents