Assembly Changed
Perform the Following Related Adjustments in the Order Listed
or Repaired
Al3 2nd Converter
Frequency Response Adjustment
Al4 Frequency Control
Display Adjustment (Fast Zero Span)

YTO Adjustment

First LO Distribution Amplifier Adjustment

Frequency Response Adjustment
10 MHz Reference Adjustment (TCXO)
Al5 RF
Calibrator Amplitude Adjustment
External Mixer Bias Adjustment

Sampling Oscillator Adjustment

Signal ID Oscillator Adjustment
External Mixer Amplitude Adjustment
Frequency Response Adjustment
600 MHz Amplitude Adjustment
Second IF Gain Adjustment
Sampling Oscillator Adjustment
Al7 CRT Driver
Display Adjustment
Display Adjustment
A19 HP-IB
No related adjustment
A21 OCXO
10 MHz Reference Adjustment (OCXO)
Table 2-1. Related Adjustments (2 of 2)
Adjustmeni
Number
10
2
7
8
10
12
11
14
6
17
15
10
18
16
6
2
2
19