Side Frame Mounting Screws - HP 8560A Service Manual

Spectrum analyzer
Table of Contents

Advertisement

6. Attach the line-switch assembly into the front frame using the captive panhead screw. Be
sure to connect the line-switch grounding lug with the screw.
7. On the top side of the analyzer, redress W3.
8. Connect W3 to A6J2. Dress W3 into the slotted opening in the deck.
9. Connect AlAlWl to A3J602.
10. Secure the power-supply cover shield to the power supply using three flathead screws. One
end of the cover fits into a slot provided in the rear-frame assembly.
11. Place W3 and the other cable assemblies between the CRT assembly and the power
supply cover so the W9 wires are below the surface of the power-supply cover.
12. Fold up the A2, A3, A4, and A5 assemblies into the analyzer as described in Procedure 5,
"A2, A3, A4, and A5 Assemblies Replacement," steps 5 through 10.
13. Fold up Al4 and Al5 assemblies as described in Procedure 9, "Al4 and Al5 Assemblies
Replacement," steps 3 through 5.
14. Replace the analyzer's cover assembly.
15. Connect the line-power cord and switch the analyzer's power on. If the analyzer does not
operate properly, turn off the analyzer power, disconnect the line cord, and recheck the
analyzer.
Figure 3-27. Side Frame Mounting Screws
Procedure 14. W3 Line Switch Cable
Assembly Replacement 3-49