9. Tracking Generator Power Level Adjustments
10. Set the TRK GEN RF POWER to
receiver.
11. Proceed with steps 12 through 14 only if the power level noted in the previous step was
outside the range of -10 dBm +/-.23 dB.
12. With the TRK GEN RF POWER set to
-10 dBm +/-.l dB reading on the measuring receiver. Refer to Figure 2-12 for adjustment
location.
13. Set the TRK GEN RF POWER to 0 dBm. Adjust AlOR 0 dB ADJ for a 0 dBm
14. Repeat steps 12 and 13 until the output power level is within the tolerances indicated at
both the -10 dBm and 0 dBm TRK GEN RF POWER settings. Adjust -10 dB ADJ
only with the TRK GEN RF POWER set to -10 dBm and adjust 0 dB ADJ only with
the TRK GEN RF POWER set to 0 dBm.
2-44 Adjustment Procedures