5. Connect A18Wl to A17J5.
6. Snap CRT cable W9 onto the end of the CRT assembly.
7. Fully seat the front frame and CRT assemblies into the analyzer.
8. Secure the front frame to the analyzer's side frames, using three flathead screws per side.
See Figure 3-4.
9. Retighten the four screws securing the CRT clamps to the deck.
10. Place W9 between the CRT assembly and the A6 Power Supply assembly top shield so
that the W9 wires are below the surface of the top shield.
11. Connect W42 to A7J3 and the front-panel 1ST LO OUTPUT connector.
analyzers:
connect W46 to the front-panel 1ST LO OUTPUT connector.
12. Use a 9/16-inch nut driver to reconnect CAL OUTPUT connector to the front panel.
13. Connect the VOLUME potentiometer and knob to the front panel.
For Option 002 analyzers:
14.
front-panel RF OUT connector.
Figure 3-6. Placing the CRT into the Front Frame
use a 5/16-inch open-end wrench to connect W47 to the
Procedure 2. Al Front Frame/Al8 CRT
For Option 002
Assembly Replacement 3-9