Resolution Bandwidth Selectivity Test; Db Bandwidth Measurement - HP 8568B Performance Tests And Adjustments Manual

Spectrum analyzer
Hide thumbs Also See for 8568B:
Table of Contents

Advertisement

5. Resolution Bandwidth Selectivity Test

2 - 1 6 P e r f o r m a n c e T e s t s
6. Read the 60 dB bandwidth for the 3 MHz resolution bandwidth
setting from the MARKER A frequency readout (see Figure 2-7)
and record the value in Table 2-9.
7. Vary spectrum analyzer settings according to Table 2-9. Measure
the 60 dB bandwidth for each resolution bandwidth setting by the
procedure of steps 4 through 6 and record the value in 'fable 2-9.
8. Record the 3 dB bandwidths from Table 2-8 in Table 2-9.
9. Calculate the bandwidth selectivity for each setting by dividing
the 60 dB bandwidth by the 3 dB bandwidth. The bandwidth
ratios should be less than the maximum values shown in
10. The 60 dB bandwidth for the 10 Hz resolution bandwidth setting
should be less than 100 Hz.
Figure 2-7. 60 dB Bandwidth Measurement