22. Comb
Generator
Adjustments
Reference
Description
3-136 Adjustments
RF Section:
The output of the Pilot First Converter is connected to the signal
input of the Second Converter. This allows the comb teeth from the
phase lock flags are disabled, using a shift key function to prevent the
instrument from "locking up" due to the phase lock loops being open.
A display line is placed on the CRT at the level to which the comb
teeth are to be adjusted. the comb teeth are adjusted for best overall
flatness and to the proper amplitude.
COMB (;ENt
IRATOR
I
\
Figure 3-83. Location of Comb Generator Adjustments