12. Time Base Adjustment (SN 2840A and Below, also 3217A05568 and Above)
Procedure
Note
Note
Note
The spectrum analyzer must be ON continuously (not in STANDBY) for
at least 72 hours immediately prior to oscillator adjustment to allow
both the temperature and frequency of the 10 MHz Quartz Crystal
Oscillator to stabilize. Adjustment should not be attempted before the
oscillator is allowed to reach its specified aging rate. Failure to allow
sufficient stabilization time could result in oscillator misadjustment.
The A27Al 10 MHz Quartz Crystal Oscillator (HP P/N 0960-0477)
will typically reach its specified aging rate again within 72 hours
after being switched off for a period of up to 24 hours. If extreme
environmental conditions were encountered during storage or
shipment (i.e. mechanical shock, temperature extremes) the oscillator
could require up to 30 days to achieve its specified aging rate.
1. Set the rear-panel FREQ REFERENCE switch on the spectrum
analyzer RF Section to INT.
The +22 Vdc STANDBY supply provides power to the heater circuit in
the A27 10 MHz Frequency Standard assembly whenever line power
is applied to the RF Section. This allows the A27 10 MHz Frequency
Standard oven to remain at thermal equilibrium, minimizing
frequency drift due to temperature variations. The OVEN COLD
message should typically appear on the spectrum analyzer display for
10 minutes or less after line power is first applied to the RF Section.
The rear-panel FREQ REFERENCE switch enables or disables the RF
Section +20 Vdc switched supply, which powers the oscillator circuits
in the A27 10 MHz Frequency Standard. This switch must be set to
INT and the spectrum analyzer must be switched ON continuously
(not in STANDBY) for at least 72 hours before adjusting the frequency
of the A27 10 MHz Frequency Standard.
2. Set the LINE switch to ON. Leave the spectrum analyzer ON
(not in STANDBY) and undisturbed for at least 48 hours to allow
the temperature and frequency of the A27 10 MHz Frequency
Standard to stabilize.
3. Press (SHIFT) TRACE B
prolongs CRT life while the spectrum analyzer is unattended. To
turn the CRT back on press m TRACE B C-HOLD) h.
4. Connect the (Cesium Beam) Frequency Standard to the Frequency
Counter's rear-panel TIMEBASE IN/OUT connector as shown in
Figure 3-52.
5. Disconnect the short jumper cable on the RF Section rear panel
from the FREQ REFERENCE INT connector. Connect this output
(FREQ REFERENCE INT) to INPUT A on the Frequency Counter.
A REF UNLOCK message should appear on the CRT display.
[
-
]
CLEAR
WRITE