HP 8568B Performance Tests And Adjustments Manual page 169

Spectrum analyzer
Hide thumbs Also See for 8568B:
Table of Contents

Advertisement

3. Preliminary Display Adjustments (SN 3004A and Above)
12. Set the oscilloscope controls as follows:
Press [CHAN]:
Channel 1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . on
amplitude scale
offset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ..25.000 0 V
Channel 4 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . on
amplitude scale
offset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ..60.000 0 V
Press (TRIG):
source . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 .
level . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25.0000 V
Press [TIME):
time scale . . . . . . . . . . . . . . . . . . . . . . . .._....__.......... 50.0 ns/div
delay . . . . . . . . . . .. . . . . . . . . . . . . .. . . . . . . . . . . . . . . 125.000 ns
Press (DISPLAY):
connect dots . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ..on
Press ISHOW].
13. Set the spectrum analyzer's front-panel INTENSITY control fully
counterclockwise, and then set the LINE switch to ON.
14. The X+ deflection and X- deflection waveforms should be
superimposed on the oscilloscope display, as shown in Figure 3-26.
If necessary, adjust AlA2R227 X POSN and AlA2R220 X GAIN
for a centered display of at least four vertical divisions. See
Figure 3-25 for the location of the adjustments.
r u n n i n g
- 1 2 5 . 0 0 0 ns
Figure 3-26. X + and X- Waveforms
15. Set the oscilloscope controls as follows:
Press
f l . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ..on
display . . . . . . . . . . . .. .. .. .. .. . .. .. .. . . . . . . . . . . . . . . . . . . .. .. on
math . . . . . . . . . . . . . . . .. . . . . . . . . . .
sensitivity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ...25.0 Vldiv
16. Three waveforms should be displayed on the oscilloscope, as
shown in Figure 3-27. The lower composite waveform represents
1 0 . 0 V/div
1
2 5 . 0 0 V
10.00
:
1 0 . 0 V/div
4
v
10.00
25.00
V
channel 1 - channel 4
Adjustments 3-55

Advertisement

Table of Contents
loading

Table of Contents