High Voltage Adjustment Setup - HP 8568B Performance Tests And Adjustments Manual

Spectrum analyzer
Hide thumbs Also See for 8568B:
Table of Contents

Advertisement

2. High-Voltage Adjustment (SN 3004A and Above)
Equipment
High-Voltage
Adjustment Procedure
Warning
Warning
Note
Digital Voltmeter (DVM) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .HP 3456A
DC High-Voltage Probe (1000: 1 divider) . . . . . . . . . . . . . . . . . . HP 34lllA
In the following procedure, it is necessary to probe voltages
which, if contacted, could cause serious personal injury. Use
a nonmetallic alignment tool when making adjustments. Be
extremely careful.
Do not attempt to measure the CRT filament voltage directly. The
filament voltage is referenced to the high-voltage cathode and
can only be measured safely with a special high-voltage true-rms
voltmeter and probe.
1. Set the spectrum analyzer's LINE switch to STANDBY.
2. Remove the top cover from the IF-Display Section and connect
the equipment as shown in Figure 3- 12.
Figure 3-12. High Voltage Adjustment Setup
3. Set the DVM to the 1OOV range, and connect the DVM to
Figure 3-13.
The accuracy of the high-voltage probe is specified for a probe
connected to a de voltmeter with 10 MO input resistance. HP 3456A
and HP 3455A digital voltmeters have a 10 MQ input resistance on the
100 V and 1000 V ranges. All measurements in this procedure should
the
be performed with
DVM manually set to the 100 V range (&OO.OOO
on the HP 3456A display).
SPECTRUM ANALYZER

Advertisement

Table of Contents
loading

Table of Contents