8. Frequency Response Test
49. For each input, subtract the lowest minimum level (greatest
negative) from the highest maximum (least negative)
measurement recorded in steps indicated. The result should not
exceed 2 dB.
SIGNAL INPUT 1
100 Hz to 500 MHz (from steps 16, 25, 42, or 48)
Overall Maximum
-Overall Minimum
Overall Deviation
SIGNAL INPUT 2
100 kHz to 1.5 GHz (from steps 12 or 31)
Overall Maximum
-Overall Minimum
Overall Deviation
50. Subtract the lowest minimum level (greatest negative) from the
highest maximum (least negative) measurement recorded in steps
indicated. The result should not exceed 3 dB.
SIGNAL INPUT 1
100 Hz to 1.5 GHz (from steps 15, 16, 25, 42, or 48)
Overall Maximum
-Overall Minimum
Overall Deviation