HP 8568B Performance Tests And Adjustments Manual page 27

Spectrum analyzer
Hide thumbs Also See for 8568B:
Table of Contents

Advertisement

Note
Spectrum analyzer center frequency readout can fall outside of
specified limits if 10 MHz frequency reference has not been calibrated
within the past year. To eliminate "frequency reference error x
tune frequency" term, substitute spectrum analyzer 10 MHz FREQ
REFERENCE rear panel output for frequency standard and repeat
test.
100 MHz
100 MHz
100 MHz
10 MHz
10 MHz
10 MHz
10 MHz
1 MHz
100 kHz
10 kHz
1. Center Frequency Readout Accuracy Test
Spectrum Analyzer
I
100
98
500
498
1000
998
100
99.8
500
499.8
1000
999.8
1500
1499.8
1000
999.98
1000
999.998
1000
999.9998
Center Readout
Min
Measured
Performance Tests 2-5
Max
102
502
1002
100.2
500.2
1000.2
1500.2
1000.02
1000.002
1000.0002

Advertisement