HP 8568B Performance Tests And Adjustments Manual page 81

Spectrum analyzer
Hide thumbs Also See for 8568B:
Table of Contents

Advertisement

13. Press MARKER Ia] and position movable marker at the lowest
the
point on
trace (see Figure 2-30). Read the MARKER A
amplitude from the display and record its absolute value.
MARKER A amplitude = p-p amplitude =
14. Divide the peak-to-peak amplitude by the slope computed in step
8 to obtain the residual FM:
p-p amplitude/filter slope = residual FM
= residual FM
The residual FM should be less than 3 Hz.
15. Press
16. Press Cm] 9 and adjust FREQ ZERO for a maximum amplitude
trace.
16. Residual FM Test
Hz