HP 8568B Performance Tests And Adjustments Manual page 299

Spectrum analyzer
Hide thumbs Also See for 8568B:
Table of Contents

Advertisement

Frequency
1
Synthesizer Calibrated
Amplitude Amplitude
Step
0 (ref)
0
- 1 0
- 2 0
- 1 0
- 2 0
- 3 0
- 3 0
- 4 0
- 4 0
- 5 0
- 5 0
- 6 0
- 6 0
- 7 0
- 7 0
- 8 0
- 8 0
- 9 0
20. Subtract the greatest negative fidelity error from the greatest
positive fidelity error for calibrated amplitude steps from -10 dB
to -70 dB. The results should be SO.6 dB.
21. Subtract the greatest negative fidelity error from the greatest
positive fidelity error for calibrated amplitude steps from -10 dB
to -90 dB. The results should be 1.1.5 dB.
2
MARKER A Amplitude (Column 2 - Column 1)
0 (ref)
Linear Fidelity
22. Key in analyzer settings as follows:
FREQUENCY SPAN_)
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 kHz
23. Set the frequency synthesizer for an output power level of + 10
24. Press SCALE LIN pushbutton. Press MARKER
[
PEAK
25. Set
(
FREQUENCY SPAN
26. Decrease frequency synthesizer output amplitude by 10 dB steps,
. noting the MARKER A amplitude and recording it in column 2 of
12. Option 857 Amplitude Fidelity Test
Fidelity Error
Cumulative Cumulative
0 to 80 dB
0 (ref)
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
to center the signal on the display.
to 0 Hz and CV'DEOBW) to 1 Hz. Press (SHIFT),
)
Error
Error
0 to 90 dB
20 kHz
Option 857 5-5