Signal Generator - HP 8559A Operation And Service Manual

Spectrum analyzer
Hide thumbs Also See for 8559A:
Table of Contents

Advertisement

5.19.
LOG AMPLIFIER LOG AND LINEAR ADJUSTMENT (SERIAL PREFIX 2208A)
REFERENCE:
A14 and A15 Schematics
DESCRIPTION:
10 dB/DIV and LIN are adjusted for correct steps and full-screen display translations.
DISPLAY
EQUIPMENT:

Signal Generator

Digital Voltmeter
Step Attenuator (10 dB/step)
Adapter, Type N Male on one end, BNC female on other end
Adapter, BNC Male on one end, SMA Male on other end
PROCEDURE:
1. Set spectrum analyzer controls as follows:
FREQUENCY BAND GHz
FREQ SPM/DIV
RESOLUTION BW
Amplitude Scale
SWEEP TIME/DIV
SWEEP TRIGGER
ADJUSTMENTS
DIGITAL VOLTMETER
FIGURE
LOG AMPLIFIER LOG AND LINEAR ADJUSTMENTTESTSETUP
.........................................................
........................................................
..................................................
..........................................................
.......................................................
..............................................................
........................................................
.....................................................
SIGNAL GENERATOR
ATTENUATOR
HP 355D, Option H82
.................
HP 1250-0780
....................
HP 1250-0831
F
UTPUT
HP 8640B
HP 3490A
300 kHz
-50
LIN
AUTO
FREE RUN