3 Operation
If the concentration detected is overrange, the first line of the display blinks
continuously.
3.8

Programming

CAUTION: The programming functions of the Set Range and
Curve Algorithm screens are configured at the facto-
ry to the users application specification. These func-
tions should only be reprogrammed by trained,
qualified personnel.
To program, you must:
1. Enter the password, if you are using the analyzer's password
protection capability.
2. Connect a computer or computer terminal capable of sending an
RS-232 signal to the analyzer RS-232 connector. (See chapter 2
Installation for details). Send the rp command to the analyzer.
3. Press the System button to start the System function.
Use the < > arrow keys to blink
Select MORE and press ENTER one more time
Now you will be able to select the APPLICATION and ALGORITHM
set-up functions.

3.8.1 The Set Range Screen

The Set Range screen allows reprogramming of the three analysis ranges
and the calibration range (background gas, low end of range, high end of range,
and % or ppm units). Original programming is usually done at the factory ac-
cording to the customer's application. It must be done through the RS-232 port
using a computer running a terminal emulation program.
Note: It is important to distinguish between this System program-
ming subfunction and the Range button function, which is an
3-26
Part I
, then press Enter .
Teledyne Analytical Instruments
Model 6000A