3 Operation
Before starting linearization, perform a standard calibration. See section
4.4. To enter data:
From the System Functions screen—
1. Use < > to select
2. Select and Enter
3. Enter
The Auto Linearize Mode data entry screen appears.
5. Use the ∆∇ keys to set the proper value of calibration gas, and
Enter . Repeat this step for each cal-point number as it appears in
the
6. Repeat step 5 for each of the special calibration gases, from the
lowest to the highest concentrations. Press Escape when done.
To end the session, send:
st<enter>
st<enter>
to the analyzer from the computer.
3-30
Part I
, and Enter .
.
from the Calibration Mode Select screen.