Teledyne 6000A Operating Instructions Manual page 20

Photometric analyzer
Table of Contents

Advertisement

2 Installation
Table 2-2: Analog Range ID Output - Example
Range
LO
MED
HI
CAL (0-25%)
Alarm Relays:
There are three alarm-circuit connectors on the alarm relays block
(under RELAY OUTPUTS) for making connections to internal alarm relay
contacts. Each provides a set of Form C contacts for each type of alarm.
Each has both normally open and normally closed contact connections. The
contact connections are indicated by diagrams on the rear panel. They are
capable of switching up to 3 ampers at 250 V AC into a resistive load
(Figure 2-6).
The connectors are:
Threshold Alarm 1:
2-6: Part I
Voltage (V)
0.25
0.50
0.75
1.00
Figure 2-6: Types of Relay Contacts
• Can be configured as high (actuates when
concentration is above threshold), or low
(actuates when concentration is below thresh old).
• Can be configured as fail-safe or non-fail-safe.
Teledyne Analytical Instruments
Current (mA)
8
12
16
20
Model 6000A