Teledyne 6000A Operating Instructions Manual page 45

Photometric analyzer
Table of Contents

Advertisement

Photometric Analyzer
4. Are either of the alarms to be defeated?
If you are using password protection, you will need to enter your password
to access the alarm functions. Follow the instructions in section 3.3.3 to enter
your password. Once you have clearance to proceed, enter the Alarm function.
Press the Alarm button on the front panel to enter the Alarm function.
Use the ∆ ∆ ∆ ∆ ∆ ∇ keys to choose between % or ppm units. Then press Enter to
move to the next screen.
Five parameters can be changed on this screen:
to non-alarm conditions. This mode requires an alarm to be
recognized before it can be reset. In the non-latching mode, the
alarm status will terminate when process conditions revert to non-
alarm conditions.
The defeat alarm mode is incorporated into the alarm circuit so
that maintenance can be performed under conditions which
would normally activate the alarms.
The defeat function can also be used to reset a latched alarm.
(See procedures, below.)
Value of the alarm setpoint, AL1: ####
Out-of-range direction, HI or LO
Defeated? Dft:Y/N (Yes/No)
Failsafe? Fs:Y/N (Yes/No)
Latching? Ltch:Y/N (Yes/No).
To define the setpoint, use the < > arrow keys to move the
blinking over to AL1: ####. Then use the ∆∇ arrow keys to
change the number. Holding down the key speeds up the
incrementing or decrementing.
To set the other parameters use the < > arrow keys to move the
blinking over to the desired parameter. Then use the ∆∇ arrow
keys to change the parameter.
Once the parameters for alarm 1 have been set, press Alarms
again, and repeat this procedure for alarm 2 (AL2).
To reset a latched alarm, go to
times or ∇ two times. (Toggle it to and then back to )
Teledyne Analytical Instruments
Operation 3
and then press either ∆ two
Part I
3-21

Advertisement

Table of Contents
loading

Table of Contents