System Self Diagnostic Test - Teledyne 6000A Operating Instructions Manual

Photometric analyzer
Table of Contents

Advertisement

4 Maintenance
4. Reassemble Housing as shown in Figure 4-1.
American Fuses
4.2

System Self Diagnostic Test

1. Press the System button to enter the system mode.
2. Use the < > arrow keys to move to
3. Use the < > arrow keys to move to
The following failure codes apply:
Table 4-1: Self Test Failure Codes
4-2: Part I
Figure 4-2: Installing Fuses
Power
0
OK
1
5 V Failure
2
15 V Failure
3
Both Failed
Analog
0
OK
1
DAC A (0–1 V Concentration)
2
DAC B (0–1 V Range ID)
3
Both Failed
Preamp
0
OK
1
Zero too high
2
Amplifier output doesn't match test input
3
Both Failed
European Fuses
, and press Enter .
, and press Enter .