Span Gas; A Recommended 2-Year Spare Parts List - Teledyne 212R Instruction Manual

Thermal conductivity analyzer
Table of Contents

Advertisement

Thermal Conductivity Analyzer
7.4.3
Containing an equivalent impurity. Again, the span gas must be procured
from a supplier who will certify its composition. The range switch must be in
range 3 when the span gas is introduced.
IMPORTANT: The accuracy of the analysis is directly related to the
users knowledge of the span gas composition.
Reference and Sample Flowrates 0.3 SCFH
Bypass Flowrate 3.0 SCFH
NOTE:
Cylinder and sample stream pressure settings must
be adjusted in order to maintain the above flowrates
when switching from one sample to another.
RECOMMENDED SPARE PARTS LIST
QUANTITY PART NO.
1
A-6251
1
A-6981
1
F-45
1
F-22
A minimum charge applies to all spare part orders.
IMPORTANT: Orders for replacement parts should include the part
number, the model and serial number of the analyzer in which they are to be
used.
26

Span Gas.

DESCRIPTION
Measuring Cell Block Assembly
Proportional Temperature Control
Sample Flowmeter Tube
Sample Flowmeter Bypass Tube
Teledyne Analytical Instruments
Model 212R

Advertisement

Table of Contents
loading