4-8.

Instrument Setup

The softkeys in the instrument setup menu (accessed by pressing "Instmt Setup" softkey
in the setup menu) are shown below.
PREV
MENU
The list below describes sub-menus accessed by each softkey and tells you where you can
find further information in the manuals.
Format EEPROM: Opens a menu that lets you restore all or part of the data in the
EEPROM to factory defaults. The use of this menu is described next.
Spec Format Setup: Opens a menu that lets you select the confidence level of the
calibration specifications and calibration interval.
Set Intrnl Clock: Accesses softkeys that allow you to check and set the date and time
for the clock/calendar. This chapter describes how to set the time and date.
Boost Amp Types: Selects the auxiliary amplifier model number for current and
voltage boost. This chapter describes how to select the boost amp type.
Remote Port Setup: Opens setup menus for the IEEE-488 instrument control port and
the RS-232C serial interface port. Chapter 5 describes how to set up the IEEE-488
interface. Chapter 6 describes how to set up the serial interface.
4-9.

Format EEPROM Menu

Use with extreme care. The format EEPROM menu presents
softkeys that permanently erase calibration constants. Pressing
"ALL" or "CAL" invalidates the state of calibration of the
5700A/5720A Series II Calibrator.
Pressing "Format EEPROM" in the instrument setup menu opens the following menu:
PREV
MENU
Caution
Front Panel Operation
The Setup Menu
4
4-7