Advertisement

5700A/5720A Series II
Operators Manual
Range
99 % Confidence Level
220 µA
2.2 mA
22 mA
[1]
220 mA
[1]
2.2 A
5725A Amplifier:
11 A
95 % Confidence Level
220 µA
2.2 mA
22 mA
[1]
220 mA
[1]
2.2 A
5725A Amplifier:
11 A
Note:
Maximum output from the calibrator's terminals is 2.2 A. Uncertainty specifications for 220 mA and 2.2 mA ranges are increased
by a factor of 1.3 when supplied through 5725A terminals.
Specifications are otherwise identical for all output locations.
1.
Add to uncertainty specifications:
±200 x I
±10 x I
For fields strengths >0.4 V/m but ≤3 V/m, add 1 % of range.
2.
1-34
5700A Series II DC Current Specifications
±5 °C from calibration temperature
Resolution
24 Hours
nA
0.1
45 + 10
1
45 + 10
10
45 + 100
µA
0.1
55 + 1
1
75 + 30
10
330 + 470
340 + 480
nA
0.1
35 + 8
1
35 + 8
10
35 + 80
µA
0.1
45 + 0.8
1
60 + 25
10
330 + 470
340 + 480
2
ppm for >100 mA on 220 mA range
2
ppm for >1 A on 2.2 A range
Absolute Uncertainty
90 Days
180 Days
± (ppm output + nA)
50 + 10
55 + 10
50 + 10
55 + 10
50 + 100
55 + 100
60 + 100
± (ppm output + µA)
60 +
65 + 1
80 + 30
90 + 30
350 + 480
360 + 480
± (ppm output + nA)
40 + 8
45 + 8
40 + 8
45 + 8
40 + 80
45 + 80
± (ppm output + µA)
50 + 0.8
55 + 0.8
65 + 25
75 + 25
350 + 480
360 + 480
Relative Uncertainty ±1 °C
[2]
1 Year
24 Hours
60 + 10
24 + 2
60 + 10
24 + 5
24 + 50
70 + 1
26 + 0.3
95 + 30
40 + 7
100 + 130
50 + 8
20 + 1.6
50 + 8
20 + 4
50 + 80
20 + 40
60 + 0.8
22 + 0.25
80 + 25
35 + 6
100 + 130
90 Days
26 + 2
26 + 5
26 + 50
30 + 0.3
45 + 7
110 + 130
22 + 1.6
22 + 4
22 + 40
25 + 0.25
40 + 6
110 + 130