Powering On The Calibrator; Warm Up - Fluke 5700A Series II Operator's Manual

Multi-function calibrator
Hide thumbs Also See for 5700A Series II:
Table of Contents

Advertisement

5700A/5720A Series II
Operators Manual

4-2. Powering on the Calibrator

To avoid electric shock, make sure the calibrator is safely
grounded as described in Chapter 2.
Before turning the calibrator on, make sure that the line voltage
selection switches are set properly for your line voltage. Refer
to Figure 2-3 or to the line voltage switch label and check the
line voltage now if you have not already done so.
When you power on the calibrator, the Control Display appears as shown below and then
displays a series of tests, as the calibrator performs self-tests. If a self test fails, a prompt
on the Control Display identifies the failed test and lets you choose whether to continue
self testing or stop and run self diagnostics. For information about self-diagnostics, refer
to Chapter 3 of the 5700A/5720A Series II Service Manual.
After the calibrator passes its self-tests, it powers up to 0V dc in standby, configured for
internal sensing and internal guard, ready to accept an entry from the keypad. The
Control Display now appears as shown below:

4-3. Warm Up

The calibrator must always be warmed up, to allow the environmentally controlled
components inside to stabilize, and to ensure that it meets or exceeds the specifications
listed in Chapter 1. Sufficient warmup times are as follows:
4-4
PREV
MENU
PREV
MENU
If the calibrator has been powered off for one hour or more, allow at least 30 minutes
of warmup time.
If the calibrator has been powered off for a length of time less than one hour, allow it
to warm up for at least twice the length of time it was turned off. For example, if it
has been turned off for 10 minutes, allow at least 20 minutes of warmup time.
WWarning
Caution

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

5720a series ii5720a series ii

Table of Contents