Generating A Printout - Fluke 5700A Series II Operator's Manual

Multi-function calibrator
Hide thumbs Also See for 5700A Series II:
Table of Contents

Advertisement

5700A/5720A Series II
Operators Manual
6-7.

Generating a Printout

Follow the procedure below to generate a printout of calibration shift result. Make sure
the serial interface is configured for compatibility with your peripheral device and is
attached with an RS-232C cable. The output will be paginated for a 66-line page.
1. Press the "Setup Menus" softkey; then press the "Cal" softkey. The calibration menu
2. Press the "Cal Reports" softkey. The calibration reports menu appears as shown.
3. Select one of the following softkeys to generate the printout you prefer:
If you have difficulty generating the printout, make sure the correct cable is attached to
the peripheral device, and the interface parameters are set to the correct values. If the
input buffer of the peripheral device fills, and the device sends the calibrator a stall
command, the calibrator will stop sending data and will start a timer. If the timeout period
expires, printing is aborted and the following message appears on the display:
To remove the message, press any key.
The output is paginated with Line Feed characters to format a 66-line page.
6-8
appears as shown below.
PREV
MENU
PREV
MENU
To print the main output calibration shift results press the "Print Cal Shifts"
softkey (see Figure 6-2).
To print the calibration check shift results, press the "Print Check Shifts" softkey.
To print raw calibration constant data as shown in Figure, press the "Print Raw
Data" softkey.
PREV
MENU