SYST:ERR?
Description: This command queries the Calibrator's error queue to retrieve the first
Query:
SYST:REM
Description: This command sets the Calibrator in remote mode for remote control
It is not possible to send or receive data over the RS-232 or Ethernet
ports when the calibrator is not in remote mode.
SYST:RWL
Description: This command sets the Calibrator in remote mode for remote control
SYST:LOC
Description: This command returns the calibrator to the local mode. This command

Using the STATus Subsystem

This subsystem is used to enable bits in the Operation and Questionable Event
registers. The Operation and Questionable Event, Enable and Condition registers can
be interrogated to determine their state. Table 5-6 give a summary of the Calibrator's
STATUS commands.
STAT:OPER:EVEN?
Description: This query command returns the content of Operational Data Event
STAT:OPER:ENAB? <DNPD>
Description: This command enables bits in the Operational Data Enable register.
Parameters: <DNPD>
5-40
SAF:TIME?
error in the queue. The error queue stores errors as first-in first-out
(FIFO). The error is deleted after being read. In the case of error queue
overflow, the last error in the queue is –350,"Queue overflow". The
earliest errors remain in the queue, while the most recent error is
discarded.
SYST:ERR?
through the RS-232 or Ethernet ports. All keys on the front panel
except the LOCAL key are disabled.
Note
through the RS-232 or Ethernet ports. All keys on the Calibrator's
front panel except the LOCAL key, are disabled.
is used only when controlling the Calibrator through the RS-232 or
Ethernet ports.
register. It is a decimal value which corresponds to the binary-
weighted sum of all bits set in the register. Register is cleared after
this query.
Selected bits are summarized at bit 7 (OSS) of the IEEE 488.2 Status
Byte register.
Returns the time in the Calibrator.
Returns first error in the error queue as an
error code followed by an error message.
Represents the register bits as a decimal
number.