Table 3-5 lists the error messages and possible corrective action you can take to attempt
to clear the problem.
Error Message
Warm start
Cold start
Value too large
Value too small
Negative value
Temperature too high
Test voltage unstable
Test voltage too high
Impedance to high
Overload voltage
output!
High temperature!
FBK error!
Interface error!
Bad command!
Bad communication!
Over range!
Bad calib. code!
Time warm up!
Internal RxD timeout!
Internal
communication!
Calibrator is not ready! Internal error
3-16
Table 3-5. Error Messages
Description
Start after internal reset
Start after switching on
Maximum value is out of limit
Minimum value is out of limit
Negative value instead of positive
was entered
Calibrator power stage
overheated
Unstable test voltage
Power line voltage is over 265 V
Residual loop/line impedance is
over 2
AC/DC calibrator output
overloaded
Internal temperature too high
Internal error
GPIB communication error
Bad command of GPIB
GPIB communication error
Exceeded the range limits via
GPIB
Bad calibration code
Attempt to start internal calibration
before warm up
Internal error
Internal error
Corrective Action
Attempt to set value over possible range. Set
correct value.
Attempt to set value under possible range. Set
correct value.
Enter correct polarity.
Residual loop or line impedance compensator
cannot be activated.
Reduce the load at the Calibrator's output.
Output stages are overloaded. Do not use
ranges 200 V, 1000 V, or 20 A for at least 10
minutes. Check if the ventilation holes are clear.
Turn off the calibrator and turn it on again.
Wrong data format on GPIB.
Unknown command on GPIB.
Listener not connected to the GPIB. Check
correct connection of GPIB cable.
Value out of range was set via GPIB. Set
correct value.
Wrong calibration code was entered, calibration
cannot start. Enter correct calibration code.
Attempt to start calibration before 60 minutes
warm up period. Wait 60 minutes before
starting calibration.
Internal error of the calibrator. Turn the
calibrator off, wait 5 seconds, then turn it on. If
the error appears again, contact Fluke.
Internal error of the calibrator. Turn the
calibrator off, wait 5 seconds, then turn it on. If
the error appears again, contact Fluke.
Internal error of the calibrator. Turn the
calibrator off, wait 5 seconds, then turn it on. If
the error appears again, contact Fluke.