VDE: Equivalent-Patient-Leakage-Current Test

The VDE:Equivalent-Patient-Leakage-Current test measures the current that flows
between the applied-parts terminals of the DUT to the PE, Mains terminals (L1 and L2),
and exposed conductive parts of the DUT (connected to RED signal connection) in
response to an isolated AC voltage source with a magnitude of at least 110 % of Mains
voltage. Figure 5-10 shows the electrical connections between the ESA601 and the DUT.
DUT power is off. The table below the diagram indicates the position of switches in the
diagram during the test.
This test is performed with both
using the
L1
MAINS
L2
TRANSFORMER
5-26
.
.
./
/
M
A
P
VDE EQUIV
EARTH
OPEN
S3
PE
1:1.1
ISOLATION
Figure 5-10. VDE: Equivlent-Patient-Leakage-Current Test Configuration
DIAGRAM
REFERENCE
S3
and
NORMAL
REVERSE
(
/
/
INSUL
ISO
VDE EQUIV
INSUL
DEVICE UNDER TEST
DUT_L1
DUT_L2
DUT_PE
CONDUCTIVE PART
RED TERMINAL
SWITCH
ESA601
NAME
EARTH
CLOSED
isolated voltage supply polarity
)
.
SWITCH
APPLIED
PART
TEST LEAD
+
MD
A
ACTION
FE
ayr14f.eps