Applied-Parts-Insulation-Resistance Test - Fluke ESA601 Operator's Manual

Electrical safety analyzer
Table of Contents

Advertisement

Applied-Parts-Insulation-Resistance Test

The Applied-Parts-Insulation-Resistance test measures the insulation resistance (at 500 V
DC) between all applied-parts connections and the DUT'S Protective Earth. Figure 5-3
shows the electrical connections between the ESA601 and the DUT. The table below the
diagram indicates the position of switches in the diagram during the test. Voltage is
available only when momentary
pressed in the normal position. DUT power is off.
+
500V
_
DC
MD
MEG
OHMS
Figure 5-3. Applied-Parts-Insulation-Resistance Test Configuration
The Applied-Parts-Insulation-Resistance test is applicable only to DUTs
with applied parts.
Action
1. Position the
2. Connect the DUT applied parts to the applied-parts jacks on the ESA601. Connect
the DUT power cord to the test receptacle on the ESA601.
3. Select
on the
ALL
4. Depress and hold down the
in the
NORMAL
5. After the meter settles, release the switch to hold and display a reading.
6. Press the
PRINT SWITCH
.
.
./
M
A
P
VDE
DUT_L1
DUT_L2
DUT_PE
EARTH
OPEN
S3
PE
SWITCH
DIAGRAM
ESA601
REFERENCE
NAME
S3
EARTH
Note
-
FUNCTION
SELECTION KNOB
selector knob.
APPLIED PARTS
.
.
./
M
A
P
VDE EQUIV
position.
to print the reading.
Performing Electrical-Safety Tests
/
(
/
/
INSUL
ISO
VDE EQUIV
DEVICE UNDER TEST
APPLIED
PART
CONDUCTIVE PART
ACTION
OPEN
to
Ω (
AP INSUL M
LEAD INSUL M
/
(
INSUL SWITCH
ISO
Local Control
) switch is
INSUL
FE
ayr07f.eps
Ω).
/
/
)
VDE EQUIV
INSUL
5
5-13

Hide quick links: