General; Environmental Characteristics; Certifications; Ce Declaration Of Conformity - LeCroy SDA Operator's Manual

Serial data analyzer
Table of Contents

Advertisement

General

Auto Calibration: Ensures specified DC and timing accuracy is maintained for 1 year
minimum.
Power Requirements: 100 to 120 VAC at 50/60/400 Hz; 200 to 240 VAC at 50/60 Hz; Power
consumption: 650 VA, 650 W max. (SDA 3010/4000A/5000A/6000A); 800 VA, 800 W max.
(SDA 4020/6020/9000/11000/13000/18000)

Environmental Characteristics

Temperature
Operating: 5 to 40 °C
Storage (non-operating): -20 to +60 °C
Humidity
Operating: Maximum relative humidity 80% for temperatures up to 31 °C decreasing linearly
to 50% relative humidity at 40 °C
Storage (non-operating): 5 to 95% RH (non-condensing) as tested per MIL-PRF-28800F
Altitude
Operating: Up to 2,000 m
Storage (non-operating): 12,192 m (40,000 ft)

Certifications

CE Compliant, UL and cUL Listed

CE Declaration of Conformity

The oscilloscope meets requirements of EMC Directive 89/336/EEC for Electromagnetic
Compatibility and Low Voltage Directive 73/23/EEC for Product Safety.
EMC Directive:
Electromagnetic Emissions:
* To conform to Radiated Emissions standard, use properly shielded cables on all I/O terminals.
SDA-OM-E Rev H
EN 61326/A3:2003
EMC requirements for electrical equipment for measurement,
control, and laboratory use.
EN 55011/A2:2002, Class A Radiated and conducted
emissions (Class A)*
EN 61000-3-2/A2:2005 Harmonic Current Emissions (Class A)
EN 61000-3-3/A2:2005 Voltage Fluctuations and Flickers
(Pst = 1)