Connecting Cables; Generator Detail - Rear Views; Rocker-Style Power Switch/Circuit Breaker - Dukane iQ Series HP User Manual

Ultrasonic hand held systems
Table of Contents

Advertisement

iQ Series,
Ultrasonic Hand Held Systems User's Manual
Connecting Cables - Quick
Start Guide
Complete the basic connections as shown below:
AC Line Input
HAND PROBE Connector
Grounding Stud
AC Power Cord Connection
Step 1. Attach the female end of the power cord
(200/240V only) to the generator's power inlet
connector -
in Figure 3-1.
A
(The 100/120V model's power cord is permanently
attached to the unit.)
Step 2. Attach the hand probe's cable connector to the
generator's HAND PROBE connection. -
Figure 3-1. Secure the connector to the system us-
ing the two jack screws attached to the connector
hood.
Step 3. Ground the generator chassis with a user–supplied
14-Gauge wire. Attach one end to the grounding
stud -
in Figure 3-1. Attach the other end to the
C
nearest grounded metal pipe or equal earth ground.
Step 4. Attach the male end of the power cord to a suit-
able line receptacle.
Optional Connections - See Page 14 for information
about the rear panel OUTPUTS connector.
Figure 3-1 Generator Detail - Rear Views
Page  1 
Power Cord Strain Relief
HAND PROBE
OUTPUTS
C
REAR VIEW
100/12O VOLT MODEL
AC Power Inlet
Depending on your generator model, line
voltage required for the generator is either
100-120 VAC at 50/60 Hertz or
200-240 VAC at 50/60 Hertz.
The unit has a power switch, and is pow-
ered ON whenever the AC line power is
live and the switch is in the ON position as
shown in Figure 3-2 below.
in
B
Push OFF
Figure 3-2 Rocker-style Power Switch/Circuit Breaker
B
A
Optional
Connections
NOTE
HAND PROBE
C
OUTPUTS
REAR VIEW
200/24O VOLT MODEL
Dukane Manual Part No. 403-577-01
Push ON