General Characteristics
Table 6-12 General characteristics
Power Supply
Power Consumption
Input Power Option
Fuse
• Terminal:
• Power line (according to country of destination):
Display
Operating Environment
U3402A User's and Service Guide
•
100 V/120 V/220 V/240 V ± 10%
•
AC line frequency 50 Hz to 60Hz
•
16 VA maximum
•
Manual ranging (100 VAC to 240 VAC ±10%)
•
1.25 A, 500 V FB fuse
•
15 A, 600 V FB fuse (internal)
•
0.25 A, 250 V SB fuse OR
•
0125 A, 250 V SB fuse
•
Highly visible vacuum fluorescent display (VFD)
•
Operating temperature from 0 °C to + 50 °C
•
Relative humidity up to 80% at 28 °C RH (non-condensing)
•
Altitude up to 2000 meters
•
Pollution degree 2
•
For indoor use only
Specifications and Characteristics
6
107