Dc Specifications; Table 6-1 Dc Resolution, Full Scale Reading, And Accuracy - Agilent Technologies U3402A User's Manual And Service Manual

51/2 digit dual display multimeter
Hide thumbs Also See for U3402A:
Table of Contents

Advertisement

6
Specifications and Characteristics

DC Specifications

Table 6-1 DC resolution, full scale reading, and accuracy [±(% of reading + count)]
Function
Rate
Slow
Voltage
Med
Fast
90
Range
Resolution
120.000 mV
0.001 mV
1.20000 V
0.00001 V
12.0000 V
0.0001 V
120.000 V
0.001 V
1000.00 V
0.01 V
400.00 mV
0.01 mV
4.0000 V
0.0001 V
40.000 V
0.001 V
400.00 V
0.01 V
1000.0 V
0.1 V
400.0 mV
0.1 mV
4.000 V
0.001 V
40.00 V
0.01 V
400.0 V
0.1 V
1000 V
1 V
Accuracy
Maximum
(One year; 23ºC
reading
± 5ºC)
119.999
±0.012% + 8
1.19999
±0.012% + 5
11.9999
±0.012% + 5
119.999
±0.012% + 5
[3]
1000.00
±0.012% + 5
399.99
±0.012% + 5
3.9999
±0.012% + 5
39.999
±0.012% + 5
399.99
±0.012% + 5
[3]
1000.0
±0.012% + 5
399.9
±0.012% + 2
3.999
±0.012% + 2
39.99
±0.012% + 2
399.9
±0.012% + 2
[3]
1000
±0.012% + 2
Typical input
Burden voltage
[1]
impedance
[2]
10.0 MW
10.0 MW
11.1 MW
10.1 MW
10.0 MW
10.0 MW
11.1 MW
10.1 MW
10.0 MW
10.0 MW
10.0 MW
11.1 MW
10.1 MW
10.0 MW
10.0 MW
U3402A User's and Service Guide