Table 4-2 DC voltage verification test

Set the calibrator output to 0 V before disconnecting it from the multimeter
C A U T I O N

input terminals.

66
Function
Reading rate
DC voltage
Slow
Medium
Slow
Medium
Input
Range
0.000 V
120 mV
0.00000 V
1.2 V
0.0000 V
12 V
0.000 V
120 V
0.00 V
1000 V
0.00 V
400 mV
0.0000 V
4 V
0.000 V
40 V
0.00 V
400 V
0.0 V
1000 V
100.000 mV
120 mV
1.00000 V
1.2 V
10.0000 V
12 V
100.000 V
120 V
1000.00 V
1000 V
360.00 mV
400 mV
3.6000 V
4 V
36.000 V
40 V
360.00 V
400 V
1000.0 V
1000 V
U3402A User's and Service Guide
Error from nominal
one year
±8 µV
±50 µV
±0.5 mV
±5 mV
±50 mV
±50 µV
±500 µV
±5 mV
±50 mV
±0.5V
±0.02 mV
±0.17 mV
±1.7 mV
±17 mV
±170 mV
±90 µV
±900 µV
±9 mV
±90 mV
±0.6 V