14.2.1 Gravity Acceleration Correction

Step1
Turn on the display.
Open the rear cover of the display unit.
Inside the display unit, press and hold the
switch to enter the calibration mode.
CAL
Then
Step2
Press the
acceleration correction mode.
Step3
Set your local gravity acceleration using
the following switches.
switch Selecting the value of a digit.
switch Selecting a digit.
Step4
Press the
value.
Step5
Press the
scale returns to the normal weighing
mode.

14.2.2 Preparation

Step 6 Confirm the environmental conditions as follows:
Maintain a constant temperature and a stable power supply.
Install the scale on a solid floor or bench where there is no draft,
vibration, strong magnetic fields or direct sunlight.
Consider section "4. Caution".
Step 7 Display normal weighing for at least 30 minutes to warm up the scale.
is displayed.
switch to enter the gravity
ZERO
switch to store the new
ENTER
switch again. Then the
CAL
HW-G Instruction Manual Page 41