A&D HW-10KGL Instruction Manual page 30

Hw-g series digital platform scales
Hide thumbs Also See for HW-10KGL:
Table of Contents

Advertisement

This function compares a display value with the upper limit (HI) and the lower limit (LO)
and displays these results.
It is necessary to set the "comparator function (
value (HI) and lower limit value (LO) in the function table, in advance to use this function.
It is necessary to install option OP-03 or OP-04, if you use the relay output of the
comparator.
Comparison results are displayed by indicators
weighing value
lower limit value < weighing value < upper limit value OK is displayed and output.
upper limit value < weighing value
Parameter List and Word Definition
The " near-zero " is within ±4 digits from the zero point in the unit of kg.
The "digit" is a unit of display, and is equivalent to the minimum graduation.
The "zero point" is the fundamental starting point to weighing.
Function table
Caution
The upper limit value (HI) must be equal to or greater than the lower limit value (LO).
The parameters of the upper limit value (HI) and the final value (HI) use the same memory.
The parameters of the lower limit value (LO) and the zero band (LO) use the same memory.
The upper/lower comparator function, the simple batch function and the full/dribble batch
function cannot be used at the same time because these parameters use common memory.
HW-G OZIM v1a
lower limit value
Meaning and Purpose
Having pressed the
current display value.
The scale always compares the display value.
Pressing the
switch the scale always compares the display value
SET
if not near-zero.
The scale always compares the display value if not near-zero.
When the display value becomes stable after pressing the
switch, the scale compares the display value. If the
pressed again the scale stops the comparison.
When the display value is stable the scale compares the display
value. It does not compare on an unstable condition.
When the display value becomes stable while not near-zero, after
pressing the
switch, the scale compares the display value. It
SET
does not compare on an unstable condition. If the
pressed again the scale stops the comparison.
When the display value becomes stable and not near-zero, the scale
compares the value.
HW-G Instruction Manual Page 28
~
LO is displayed and output.
HI is displayed and output.
switch, the scale will always compare the
SET
)" parameters, upper limit
SET
switch is
SET
switch is
SET