Counting Mode
In counting mode the total weight of articles is converted to a count and assumes
all articles have the same mass value.
It is necessary to store a unit mass to count articles.
Step 1 Press the
Step 2 Press the
a unit mass.
Step 3 Press the
samples. The greater the quantity of samples, the
greater the accuracy of the count. Select from
5 pieces, 10 pieces, 20 pieces, 50 pieces, 100
pieces
Step 4 Put the container onto the pan.
Press the
Step 5 Put in the number of samples selected at step 3.
Press the
the stability mark is displayed. The count is then
displayed.
Caution
When the sample is too light and it is not possible to
calculate a unit mass, the scale displays lo ut and
returns to step 3. It is necessary to increase the
number of samples.
Pressing the
ENTER
displayed, the next unit is displayed.
When the unit mass is too light to store, the scale
displays lo ut .
Step 4
Remove all items from the pan.
switch to display the unit
MODE
switch to enter the mode to store
SET
switch to select the number of
switch.
TARE
switch to store the weight after
ENTER
switch after the lo ut . is
HW-G Instruction Manual Page 21