Tyscor VS 2 Plus / VS 4
EN
Installation and operating instructions
0297

Advertisement

loading
Need help?

Need help?

Do you have a question about the Tyscor VS 2 Plus and is the answer not in the manual?

Questions and answers

Summary of Contents for Durr Dental Tyscor VS 2 Plus

  • Page 1 Tyscor VS 2 Plus / VS 4 Installation and operating instructions 0297...
  • Page 2 The current version of the installation and operating instructions is available in the Download Center: http://qr.duerrdental.com/7188100021 © 2019-2025 DÜRR DENTAL SE...
  • Page 3 Contents Contents Quick Start ....Lag time ....Eco Stop .
  • Page 4 Contents Usage 10 LEDs ......11 Operating the touch screen ..11.1 Querying messages .
  • Page 5 Product for the disinfection and cleaning These installation and operating instructions of suction units apply to: Tyscor VS 2 Plus REF: 7188200100 Wear protective gloves. Tyscor VS 4 REF: 7188100100 Disconnect all power from the unit.
  • Page 6 Important information Safety Manufacturer The unit has been developed and designed in such a way that dangers are effectively ruled out if used in accordance with the Intended Use. Fragile, handle with care Despite this, the following residual risks can remain: This way up / store and transport in an –...
  • Page 7 Important information General safety information The following groups are not permitted to operate or use a commercially operated unit: – Always comply with the specifications of all – People without the necessary experience and guidelines, laws, and other rules and regula- knowledge tions applicable at the site of operation for the –...
  • Page 8 Important information 2.10 Transport The following administrative measures are required: The original packaging provides optimum protec- – Train employees in security and data protec- tion for the unit during transportation. tion. If required, the original packaging for the unit can –...
  • Page 9 Product description Product description Overview 1 Combination suction unit 2 Mains cable with local mains plug 3 Hose LW 50, 1.5 m 4 Hose LW 50, 0.6 m 5 Waste water hose LW 20 6 Set of connection fittings 7 OroCup 7188100021L02 2502V008...
  • Page 10 Backflow preventer valve (pack of Tyscor VS 2 Plus ....7188200100 2) ......7188100082 –...
  • Page 11 Product description Technical data Electrical data 7188100100 7188200100 Rated voltage 230, 1~ 230, 1~ Mains frequency 50 / 60 50 / 60 Nominal current Active power Fuses 2x T 10.0 AH / 250 V~ (IEC 60127-2) Type of protection IP 21 Protection class Control connection electrical data Output:...
  • Page 12 Product description General data Sound pressure level in accordance with ISO 3746 Network connection LAN technology Ethernet Standard IEEE 802.3u Data rate Mbit/s Connector RJ45 Type of connection Auto MDI-X ³ CAT5 Cable type Ambient conditions during storage and transport Temperature °C -10 to +60...
  • Page 13 Product description Characteristic curves Fig. 1: Characteristic curves for 7188200100, measured in accordance with ISO 10637 Fig. 2: Characteristic curves for 7188100100, measured in accordance with ISO 10637 7188100021L02 2502V008...
  • Page 14 Product description Type plate The type plate can be found on the housing lower part. 1 Type plate Evaluation of conformity This device has been subjected to conformity acceptance testing in accordance with the cur- rent relevant European Union guidelines. This equipment conforms to all relevant requirements.
  • Page 15 The suction unit is switched on when a suction hose is removed from the hose manifold. With the optional upgrade kit, the Tyscor VS 2 Plus can be upgraded to the Tyscor VS 4 to support up to four practitioners.
  • Page 16 Product description Touch screen Separation system In the separation system the aspirated fluids and the solid particles are separated from the suction air. The waste water pump in the separation system feeds the fluid together with the finer solid parti- cles through the waste water system connection into the central waste water network.
  • Page 17 Product description will switch off automatically after a pre-defined Tyscor Tandem: Combination of several period of time (it is possible to set this up so that Tyscor suction units it can be adjusted via the monitoring software). To switch it on, lift a suction hose up out of the hose manifold or briefly disconnect the power supply from the unit.
  • Page 18 Product description 5.12 OroCup The OroCup care system is a closed dosing sys- tem for simple preparation and suctioning of dis- infectants and special cleaners. Use OroCup to clean and disinfect the suction system with all of its components and the spit- toon.
  • Page 19 Assembly The following materials must not be used: Assembly – Acrylonitrile-butadiene-styrene (ABS), – Styrene copolymer blends (e.g. SAN + PVC). Requirements Hose materials For waste connections and suction lines only The unit can be installed on the same level as the use the following hose types: surgery room or in a floor below.
  • Page 20 Assembly System components Installation type Line layout (minimum requirements) The system components listed below are Flexible – PVC data cable with required or recommended for various procedures shielded cable sheath- or for installation. ing, as used for tele- Rinsing unit communications and IT processing systems It is recommended that the suction system is...
  • Page 21 Assembly accelerator. The next time suction takes place using the large cannula, the collected fluid is transported in surges and at high speed to the suction system. This ensures automatic cleaning of the suction pipes. 7188100021L02 2502V008...
  • Page 22 Assembly Installation The actual connection can vary depending on the chosen installation option. The connection shown is only an example. Installation and routeing of hoses and pipes – Establish connections between the pipe system and the unit using the flexible hoses supplied. This will prevent vibrations from being transmitted to the pipe system.
  • Page 23 Assembly Installing the group of devi- Various mounting kits are available for installing several devices in a device network (see also "3.2 Optional items"): – Installation set for Tyscor V/VS Tandem for the combination of up to three Tyscor suction machines –...
  • Page 24 Assembly Network connection CAN bus Purpose of the network connection Purpose of the CAN bus connection The network connection is used to exchange In a grouping of up to three devices, the CAN information or control signals between the unit bus is used to enable mutual controlling and and a software installed on a computer, in order information exchange between the devices.
  • Page 25 Assembly 1 Mains cable with socket and country-specific mains plug 2 Control line 7188100021L02 2502V008...
  • Page 26 Assembly PCB (main board) electrical connections X14 X13 X15 X16 Supply voltage 230 V Control voltage output, 24 V DC, control voltage input Voltage supply for the rinsing unit, 24 V DC SD card holder (for Micro SD) CAN bus Service interface Connection for the display Supply voltage for suction motor 1...
  • Page 27 Assembly Green LED – HV voltage supply +15 V Green LED – HV voltage supply +5V Green LED – HV voltage supply +3.3 V Green LED – temperature indicator for radial blower 1, temperature OK H10 Yellow LED – line to radial blower 1, short circuit or open circuit H11 Red LED –...
  • Page 28 Assembly Commissioning Testing suction power after installation In many countries technical medical prod- ucts and electrical devices are subject to Measure negative pressure regular checks at set intervals. The owner The following procedure is recommended in must be instructed accordingly. order to measure the maximum negative pres- sure of the suction system so that it can be com- NOTICE...
  • Page 29 Assembly Setting up the group settings: Use the Settings button to switch from the display screen to the menu screen. Use the Parameters button to switch to Cluster Settings. In the selection list, select whether the device is a main unit unit or an auxiliary unit. Control strategy in group operation If required, you can change the control strategy of the device or device group.
  • Page 30 Assembly Setting the afterrun delay time Exhaust air filter Depending on the installation, the lag time of the If an exhaust air filter is used, this will need to be device can be adjusted. The lag time can only be set up in the controller.
  • Page 31 Assembly Monitoring the unit via the Network configuration network Various options are available for network configu- ration: If the device is connected to the VistaSoft Moni- tor monitoring software via the local practice net- ü Automatic configuration via DHCP (recommen- work, the software displays the device’s mes- ded).
  • Page 32 Usage Usage When using prophy powders, water-soluble Lunos Prophy Powders are recommended in order to protect the suction systems (Dürr Dental). 10 LEDs Network ORANGE display is illuminated The machine is connected to the network. Error Red display flashes – Fault/failure of a radial blower RED LED illuminates –...
  • Page 33 Usage 11.1 Querying messages 11 Operating the touch Any incidents that occur in the unit are shown in screen the Messages screen. The messages can be confirmed by tapping the navigation button. The The selected operating mode and the current messages are then displayed in the context area negative pressure are displayed on the Main of the Home screen until the incident has been...
  • Page 34 Usage 12 Disinfection and cleaning The following tasks are required for the disinfection and cleaning of the suction system: "12.1 Suctioning water" After every treatment "12.2 Disinfection" Daily in the evening after the end of treatment, With increased workloads, before the midday break and after the end of treatment "12.3 Cleaning"...
  • Page 35 Usage To pre-clean, suction up 2 litres of water with 12.1 Suctioning water the care system. After every treatment: Suction up cold water (min. 0.5 litres) with the large and small suction hoses. Do this even if only the small suction hose was actually used during treatment.
  • Page 36 Usage Using the solution: 12.3 Cleaning Clean the suction system regularly: Attach suction hoses to the care system and – Allow to act for at least 2 x per week in the suction up 1 litre of solution. morning before the start of treatment or during the midday break, reaction time 30–120 minutes –...
  • Page 37 Usage Fill the OroCup with 2 litres of cold water. Suction up 2 litres of water after the reaction time has elapsed. The instructions are also available as a video: Close the lid of the care system. Mix the solution. Attach suction hoses to the care system and suction up one litre of solution.
  • Page 38 Usage 13 Maintenance All maintenance work must be performed by a qualified expert or by one of our Service Techni- cians. Maintenance interval User maintenance work Every 3 months Check the filter at the device suction connection and clean if necessary. ❯...
  • Page 39 Usage 13.4 Intensive cleaning 13.2 Clean the filter at the device Intensive cleaning can be carried out during the suction connection. self-clean process of suction systems that are Clean the filter at the suction connection every 3 heavily contaminated with biofilms or if the suc- months.
  • Page 40 Usage Fill the OroCup with 2 litres of cold water. Suction up 2 litres of water after the reaction time has elapsed. Close the lid of the care system. Mix the solution. Attach suction hoses to the care system and suction up one litre of solution.
  • Page 41 Troubleshooting Troubleshooting 14 Tips for operators and service technicians Any repairs exceeding routine maintenance may only be carried out by qualified personnel or our service. WARNING Infection due to contaminated unit Clean and disinfect the suction before working on the unit. ❯...
  • Page 42 Troubleshooting Error Possible cause Remedy Water leaking from the Membrane valve defective Check the membrane valve at ❯ the waste water connection exhaust air connection and if necessary clean or replace. Foam in turbine due to use of Use non-foaming cleaning ❯...
  • Page 43 Troubleshooting Error Possible cause Remedy Display does not respond to Display not calibrated Calibrate the display. ❯ touches – Disconnect and reconnect the mains plug. Within 8 seconds of the device being started, start the calibration process with a long press (5 s) on the display.
  • Page 44 Troubleshooting 14.2 Fault with Tyscor Tandem Flex Error Possible cause Remedy No suction power Side channel suction unit defec- Repair or replace side channel ❯ No error message on Tyscor tive suction unit.* touchscreen. Change operation to Tyscor, ❯ Error message on VistaSoft see "Emergency mode: Monitor may appear if side Change to Tyscor".
  • Page 45 Troubleshooting 14.3 Error messages Error messages are displayed on the touch screen. If there is a network connection, the mes- sages can be forwarded to the monitoring software. If the device is not connected to the net- work, the messages can be read via a terminal client (e. g. PuTTY). Error Possible cause Remedy...
  • Page 46 Troubleshooting Error Possible cause Remedy Speed of Sepa is low Motor defective Replace the separation stage. ❯ Hall sensor PCB (main board) Replace the Hall sensor PCB ❯ defective (main board), check the mag- nets in the Sepa fan. * Centrifuge soiled or damaged Check the centrifuge and ❯...
  • Page 47 Troubleshooting Error Possible cause Remedy Power Pack overheated Insufficient ventilation or poor Check the setup conditions, ❯ set-up conditions ensure adequate ventilation. Fan on the electronics housing Remove the cover on the ❯ soiled electronics housing, clean the fan and heat sink. * Fan on electronics housing Replace the fan.
  • Page 48 Troubleshooting 15 Transporting the unit WARNING Infection due to contaminated unit Disinfect the unit before transport. ❯ Close all media connections. ❯ Wear protective equipment to avoid any risk of infection (e.g. liquid-tight protective gloves, protective goggles, face mask). Prior to disassembly, clean and disinfect the suction unit and the unit via aspiration of a suitable disinfectant approved by the manu- facturer.
  • Page 49 Appendix Appendix 16 Menu structure Setup wizard Language Operating mode Exhaust air filter Finished Start screen Operating mode, vacuum Performance, tem- perature Control strategy options Settings Access levels Operator Administrator Service Technician XXXXXX [PIN] Device information Device data Firmware Upgrade Library version PCB serial number File...
  • Page 50 Appendix Settings (con- Cluster settings Main unit Parameters tinued) Auxiliary unit 1 Auxiliary unit 2 XX s Lag time Quick Start Start / stop Eco Stop Exhaust air filter Exhaust filter pres- No exhaust air filter Control strategy Adaptive operation options Parallel operation Tyscor Tandem Flex...
  • Page 51 Appendix 17 Handover record This document confirms that a qualified handover of the medical device has taken place and that appropriate instructions have been provided for it. This must be carried out by a qualified adviser for the medical device, who will instruct you in the proper handling and operation of the medical device. Product name Order number (REF) Serial number (SN)
  • Page 52 Appendix 18 Country representatives Country Address Уповноважений представник в Україні: Приватне підприємство “Галіт” вул. 15 квітня, 6Є, с. Байківці, Тернопільський р-н, 47711, Україна тел.: 0800 502 998; UA.TR.099 +38 050 338 10 64 www.galit.te.ua; e-mail: office@galit.te.ua Виробник: Дюрр Дентал ЕсE Хьопфігхаймep Штрассе...
  • Page 56 Hersteller / Manufacturer: DÜRR DENTAL SE Höpfigheimer Str. 17 74321 Bietigheim-Bissingen Germany Fon: +49 7142 705-0 www.duerrdental.com info@duerrdental.com...