Care & Maintenance; Troubleshooting - DeVilbiss iFILL PD1000A Series Instruction Manual

Oxygen cylinder
Hide thumbs Also See for iFILL PD1000A Series:
Table of Contents

Advertisement

Available languages

Available languages

E N G L I S H
CARE & MAINTENANCE
The DeVilbiss iFill Oxygen Cylinders should be kept clean and free from moisture and dust. The cylinders should be protected
from extreme temperatures. Clean the cylinders periodically by wiping with a dry, lint-free cloth. Do not clean with a solvent
based cleaning solution. Avoid dropping the cylinder or placing it in a position where it could topple or fall since this can damage
the device. Whenever possible, use a padded carrying bag such as the Hideaway or Walkabout bags to carry the cylinder. This
will help to protect it in case of a fall. Avoid getting fluids or debris such as sand or dirt inside the oxygen cylinder and/or oxygen
connector.

TROUBLESHOOTING

Do not attempt to open the device for maintenance or repair. The device contains no
Symptoms
Oxygen is not being delivered even though
the PulseDose indicator is flashing every
time I breathe.
Use times are different from those stated
in the literature.
PulseDose will not pulse
PulseDose works fine for a couple of min-
utes, then sensitivity seems to drift and
may stop working altogether.
A-535-CYL
user-serviceable parts. Contact your oxygen provider if service is required.
1. Oxygen supply is empty.
1. PulseDose responds to your breath
rate. Your breath rate may vary, which
causes the operation time to vary.
2. Leak in system
1. Cannula is not adjusted properly.
2. Unit is not turned on.
3. Batteries discharged or not installed.
4. Mouth breathing with closed soft
palates.
5. Unit did not reset while changing bat-
teries (red light stays on).
1. Using pediatric cannula or any cannula
that restricts continuous flow capaci-
ty of 10 lpm.
CA RE & M AI N TE N AN CE / TR OU B LE SHOO TI N G
WARNING
Possible Causes
10
Remedies
1. Check contents indicator on the
device. If empty, switch cylinders.
1. PulseDose is operating correctly.
2. Contact your Sunrise Medical
provider.
1. Check all cannula connections to
make sure they are tight, and adjust
the cannula to fit comfortably in your
nose. Ensure tubing is not kinked.
2. Turn the rotary selector to the appro-
priate setting.
3. Install new batteries.
4. Breathe through the nose (cannula).
5. Turn the unit off and back on using
the rotary selector.
1. Replace with standard nasal cannula.

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ifill 535-cf series

Table of Contents