1. Enter Test Mode Test mode provides access to calibration, and to system and factory parameter modes. Test Mode is only accessible by opening the scale and shorting the test mode jumpers. Follow the steps below to perform Open the scale correctly. Opening the Scale housing screws 1.
Page 3
In test mode, the following types of the internal count are displayed. When the key is pressed, the internal count, power supply voltage, display check and others are cyclically shown on the panel in the following order. TEST MODE User & Dealer Parameters Internal Count disp [1.
2. Changing Parameters NOTE) The setting method for each parameter mode is the same. Once in parameter mode use the key to select the desired parameter, then use the keys to change the set value to the desired value. The arrows above these keys indicate their function.
Page 5
Function Value Description Communication device Bluetooth (Yamato standard protocol) (not available) RS232C (Yamato standard protocol) (not available) Bluetooth wireless printer(for designated model) No communication (not available) Send contents Net weight Net weight, Tare weight, Gross weight Net weight, Tare weight...
Page 6
Dealer Parameters Enter test mode. With the display showing internal counts, press to enter User + Dealer parameter mode. Function Value Description Decimal point Type of decimal point Comma Packing weighing Invalid function Valid Invalid Checkweighing function Valid Invalid Grading function Valid Invalid Counting function...
System Parameters With the display showing internal counts, press to enter calibration mode, then press to enter System parameter mode. Function Value Description No compensation 1~29: * Japa only Gravity compensation 30~210: (Acceleration of Gravity (m/s2) - 9.7600) x 10,000 ÷ 5 + 30 Setting range;...
Page 8
Entered by key operation & TEST switch, User parameter is invalid Entered by TEST switch only (key operation is invalid) Entered by TEST switch only, User parameter is invalid UDS-211W Model UDS-211Be PPC-200W-II Reserved * Do not change the settings Reserved * Do not change the settings Reserved...
Page 9
Function Value Description Board sensitivity 0~255: Automatically set at board sensitivity adjustment (Prohibit to change) adjustment 1, zero Board sensitivity 0~255: Automatically set at board sensitivity adjustment (Prohibit to change) adjustment 2, zero Board sensitivity 0~255: Automatically set at board sensitivity adjustment (Prohibit to change) adjustment 3, zero Not use default setting 1~6:...
3. Default setting List of initial parameter values before shipment by scale type/weighing capacity The following table shows the initial setup values before shipment by type/weighing capacity. If you replace the CPU board, make sure to initialize the board using a setup value (Initialize from 12 to 16 in Default setting mode (enter [B.
4. Calibrate the scale Calibration can be done with pound or kilogram weights. Pound weights are the default setup, but this can be changed through System parameter 50. The adjustment corrects the hysteresis due to the load and unloading of the sensor. Therefore, follow the procedure below.
Page 16
Description Indication If the scale displays error 103 the scale is either misconfigured, incorrect weights were use for the calibration, or the scale may be damaged. Check that the weights are one half of capacity, and full capacity.