Statement Kaixin has the final explanation right of this user’s manual. Kaixin was considered responsible for the safety, reliability and performance in case of meeting all the following requirements: 1. Assembly, expansion, readjustment, improve and repair are all performed by professionals recognized by Kaixin;...
This user’s manual contains warnings for predictable dangers. Users shall also exercise care at any time to be aware of the dangers unforeseen in this manual. Kaixin shall not be liable for the damages and losses arising out of neglecting to follow the operation instructions herein described.
Warning Symbols and Definitions The following warning symbols are used in this user’s manual to indicate safety level and other important items. Please remember these symbols and understand the meaning as you read this user’s manual. These symbols convey specific meanings as detailed in the table below: Symbols &...
Page 7
Use of multiple portable socket-outlets may affect protective earth to make leakage currents exceed the safety requirements. You must use the power adapter provided by Kaixin; otherwise electric shock may result. You must adopt the power supply method provided by Kaixin, other power supply modes may result in electric shock.
Page 8
15. When the instrument works abnormally, do stop working, turn off the power and check the reason, then contacts the Kaixin Company about it. 16. It is forbidden to load and unload the probe or move the instrument in galvanic to avoid danger of safety.
Page 9
Warning: 1. Check the probe and connecting cable before and after diagnostic operation. Use of defective probe may cause electric shock. 2. Do not strike or hit the probe; using the damaged probe may cause electric shock to the patient. 3.
Chapter 2 System Configuration 2.1 Standard Configuration Main unit 1 unit 6.5MHz animal transrectal linear array probe 1 PC Power adapter 1 PC Power cable 1 PC Built-in battery 1 PC Hood 1 PC Socket screwdriver 1 PC Straps 2 PCS Probe holder 1 PC APP software...
Chapter 3 Product Introduction 3.1 Structure Composition of the Instrument KX5600G Full Digital Ultrasonic Diagnostic Instruments (Vet) are composed of main unit, probe and power adapter etc. 3.2 Components Name 3.3 Parts of the Probe Take 6.5MHz animal transrectal linear array probe for example:...
Figure: Parts name of 6.5MHz animal transrectal linear array probe Name Function To convert electric signal to ultrasonic signal based on principle of converse piezoelectric effect. The ultrasonic signal, after entering the (1) Acoustic lens human body, is reflected as echo wave and converted to electric signal again.
Page 14
Symbols Description Type B applied part Indicates the need for the user to consult the instructions for use for important cautionary information. Follow instructions for use Power switch Power supply indicator Battery charge indicator Direct current Polarity of direct current power connector USB port Manufacturer Date of manufacture...
Chapter 4 Installation and Check Warning : Heat sink at the back of main unit to allow heat release from ultrasound instrument. When main unit is working, the temperature of heat sink may be high. If used, be careful not to touch the heat sink, so as not to burn! Warning : Do not lift the probe to move the machine.
4.2.1 Ultrasonic Probe Connection Warning: Before connecting or using the probe, make sure that the probe, connecting cable and connector are in normal condition (free of cracks or drop). Use of defective probe may cause electric shock. Turn off the instrument, align insert the probe connector into the probe socket at the back of the main unit.
Warning: 1. Should be used the power adapter provided by Kaixin. If other adapters are used, it may cause electric shock. 2. Please use the power adapter in the environment specified in the user's manual.
1. The temperature, humidity and atmospheric pressure shall meet the requirements of operation condition. 2. No condensation occurs. 3. No distortion, damage or contamination on system and peripheral. Clean the parts as specified in relevant sections, if the contaminant is present. 4.
Chapter 5 Functional Operation 5.1 Startup and Shutdown In shutdown status, press key, machine starts up, power supply indicator lights. In startup status, hold down key, machine shuts down, power supply indicator goes out. Please note that when shut down the machine, the time of pressing key is a rather long than normal pressing key.
5.2.4 Characters Brightness Setting to move cursor to “Setup”; 1. Press key to enter main-menu, press direction keys 2. Press key to enter setting interface; to move the symbol “<” to point to “Font Bright”; 3. Press direction keys 4. Press direction keys to select characters brightness among 160, 192, 224 and 255;...
2. Press key to enter setting interface; to move the symbol “<” to point to “Grid Set”; 3. Press direction keys to select “00”-“08”; 4. Press direction keys “00” shows “no grid”, “01” shows “line grid”, “02” shows “dot grid”, “03”...
2. 4B image switch. Press key to enter main-menu, press direction keys to move the cursor to “4B Mode”, then press direction keys to switch display among four images; the selected image is activated and the other three are frozen. 3.
Page 23
5.4.2 Total Gain Adjustment In real-time mode, press key to enter main-menu, press direction keys to move cursor to “Gain” in the display area. Press direction key to increase image total gain and direction key to reduce total gain so as to control the total gain of the entire image. 5.4.3 Near Field Gain Adjustment In real-time mode, press key to enter main-menu, press direction keys...
Page 24
5.4.10 Edge Enhancement Adjustment In real-time mode, press key to enter main-menu, press direction keys to move cursor to “IE” in the display area and then press direction keys to gain four levels of sharpened images. 5.4.11 Local Zoom and Local Additive Color In real time mode, press key to enter main-menu, press direction keys to move cursor...
5.4.17 Image Freeze/Unfreeze In real-time mode, press key to freeze the image; in frozen status, press key to unfreeze the image. 5.5 Auto Ruler (Automatic Backfat Measure) 1. In real-time B mode, press key to enter main-menu, press direction keys move cursor to “AutoRuler”...
5. Press key to quit froze status. 5.8 One-key Storage Image Press key to freeze the image, press key, the system emits a beep, and then the current frozen image is stored in the main unit. Note: 1. One-key storage can saved the current frozen image into the inter memory of main unit;...
Page 27
Attention key to switch between “Local” and “U disk”. 1. When saving an image, press 2. When the U disk is inserted, it will be stored in the U disk by default; if the U disk is not inserted, it will be stored in the “Local” by default. 3.
1. Plug U disk; 2. Read out the image in the main unit according to Chapter 5.9.2, such as choose “005*”, “Img” character showed in the lower right corner of the screen in a highlighted state; key, the system emits a beep, and the current “005*” image is dumped to the 3.
5.11 Text Input Operation steps: to move cursor to “Note” in 1. Press key to freeze the image, press direction keys the display area; key, the cursor is located behind “ID”; 2. Press 3. Press key, the characters input menu will be shown at the bottom of the screen: Caps 0 1 2 3 4 A B C D E F G H I J K L M Shift...
Page 30
5.12.2 Delete Check Lists to move cursor to “LIST”, press 1. In the frozen status, press direction keys to enter check list interface; 2. Press key, clear all the check lists; 3. Press key to exit check list interface and press key to return to real-time status.
Chapter 6 Measurement 6.1 General Measurement 6.1.1 Distance Measurement 1. In B, B/B or B/M mode, freeze the desired image, the cursor is located in the “Meas” position of display area; 2. Press key, the measurement methods are showed in the lower left of the screen, press to choose “1.
Page 32
” on the right part of the screen automatically; A: 00000mm , V: 00000 mm 6. Press key again to quit the minor axis status; Press key to exchange the starting point and end point; 7. Press key to finish the first measurement; 8.
“----”sign or invalid value. Time: denotes cycle measured in ms (millisecond) HR: denotes heart rate measured in times/minute (times per minute) EF: denotes slope measured in mm/s (millimeter per second) The accuracy of software measurement: distance measurement ≤ Attention: 0.1mm; area measurement ≤ 1mm ;...
to choose “Clear” in display area and press Press direction keys key to clear all marks and data; Press key to unfreeze and return to real-time status. 6.2.3 Measurement of Swine’s Lean percentage Lean percentage calculation formula uses NSIF formula. to choose “OB”...
Chapter 7 APP software instructions 7.1 APP Software Introduce The APP software developed by our company needs to be used with the company's Full Digital Ultrasonic Diagnostic Instruments. Users need to install the APP software “KX56S” provided by our company on a smart phone or tablet. The APP software “KX56S”...
7.3 Use APP Software APP software only supports horizontal screen. All the icons on the APP software may be different due to the display screen of the phone or tablet; the display interface may be slightly different, the actual products shall govern. The usage for the phone and tablet is similar, this article only take Android phone as an example to introduce.
2. Clean the probe with sterile water to remove the residual chemicals after disinfection, because the residual chemicals may be harmful for humans. 3. Kaixin Company will not make any guarantee for the efficacy of disinfector. Please contact the appropriate manufacturer for details.
Page 38
(3) After finishing the rinsing, use a sterilized cloth or gauze to wipe the water on the surface of probe. Do not dry the probe by heating it. 2. High-level disinfection Please follow the disinfection method provided in this user’s manual for disinfection. (1) Before disinfection, wear sterilized gloves to prevent possible infection;...
8.2.1 Use for the rechargeable lithium ion battery pack The battery pack (model HYLB-1614) is charged within main unit or with charger (model DVU50) provided by KAIXIN Company. Service personnel or operator can replace the battery pack. Charging in the main unit Install the battery in the main unit in accordance with chapter 4.3, connect the main...
Page 40
charged in the main unit. Power off and charge, the charging indicator on the keyboard is green and blinking state; the charge indicator is green and no blinking when fully charged. Battery is fully charged for about 7 to 8 hours. ...
Page 41
Attention: 1. Do not throw the battery into water or be wet, which will lead to the battery leakage, explosion or fire; 2. Do not use or store the battery near the heat source, such as fire or heater, which will lead to the battery leakage, explosion or fire;...
The probe, battery, battery charger, power adapter and power supply cord as described in this section may be replaced by qualified medical personnel (operator). But these parts must be provided by KAIXIN or his authorized supplier. 8.4 Backup Data To conduct the data management regularly, data backup to external storage medium to protect data security.
8.6 Periodic Safety Checks To ensure the system performance and safety, it must be checked after using 1 year. When check the instrument, please consult Kaixin or its dealers, as they need to have professional technology engineers. Inspect sorts Inspect items...
Chapter 9 Storage and Transportation 1. If the instrument is stored over 3 months, take out the instrument from the packing case, connect it to power supply for 4 hours, and then disconnect the power and place it in the case again following the direction indicated by arrows on the package.
Chapter 10 Safety Classification 1. Classified according to the type of protection against electric shock: Class I EQUIPMENT + internally power EQUIPMENT 2. Classified according to the degree of protection against electric shock: Type B applied part 3. Classified according to the degree of protection against harmful ingress of water: The main unit belongs to IPX0 4.
Page 46
KAIXIN ELECTRONIC XUZHOU KAIXIN ELECTRONIC INSTRUMENT CO., LTD. Kaixin Mansion, C-01, Economic Development Zone, Xuzhou, Jiangsu, China Post Code: 221004 Tel: +86-516-87732932/87733758 Fax: +86-516-87732932/87792848 Website: http://www.kxele.com E-mail: info@kxele.com Information contained in this manual is subject to change without further notice.