KAIXIN DCU10 Full Digital Color Doppler Ultrasonic Diagnostic Instruments (Vet) User’s Manual XUZHOU KAIXIN Electronic Instrument Co., Ltd.
Page 2
The qualified service personnel who get Kaixin written authorization can repair the instrument out of warranty by themselves. But this should be agreed by Xuzhou Kaixin Electronic Instrument Co., Ltd. We will provide circuit diagrams, component part lists or other information to assist service personnel to repair those parts of our equipment that are designated by our company as repairable by service personnel.
Page 3
This user’s manual contains warnings for predictable dangers. Users shall also exercise care at any time to be aware of the dangers unforeseen in this manual. Kaixin shall not be liable for the damages and losses arising out of neglecting to follow the operation instructions herein described.
Page 4
User’s Manual V1.01 Full Digital Color Doppler Ultrasonic Diagnostic Instruments (Vet) Safety Cautions 1. Warning Symbols and Definitions The following warning symbols are used in this manual to indicate safety level and other important items. Please remember these symbols and understand the meaning as you read this user’s manual. These symbols convey specific meanings as detailed in the table below: Symbols &...
Page 5
User’s Manual V1.01 Full Digital Color Doppler Ultrasonic Diagnostic Instruments (Vet) 3. Labels Label Description Danger: 1. It may have explosion hazard if used with flammable gas. 2. Operate the probe with care. Read the probe information in the relevant manual for proper probe use.
You should familiar with the structure composition, operation of the keys and buttons, the general work flow before using this instrument. 2.1 Structure composition of the instrument DCU10 full digital color Doppler ultrasonic diagnostic instruments (Vet) are composed of main unit and probe, etc. 2.2 Main Unit Overviews Figure 2-1 Main Unit Overview 2.2.1 Left View...
User’s Manual V1.01 Full Digital Color Doppler Ultrasonic Diagnostic Instruments (Vet) 2.2.2 Rear View Figure 2-3 Rear View 2.3 Peripherals Ports View Figure 2-4 Peripherals Ports Name Function Used for connecting the video equipment generating VGA VGA port signals, such as monitor and projector. Used for connecting the debugging cable and exporting the Serial port debugging information for after-sales servicing.
User’s Manual V1.01 Full Digital Color Doppler Ultrasonic Diagnostic Instruments (Vet) 2.4 Power Supply Port Figure 2-5 Power Supply Port Name Function AC power input Used for connecting the AC power supply. port Equipotential Used for balancing the protective earth potentials between Terminal this instrument and other electrical equipment.
Page 14
User’s Manual V1.01 Full Digital Color Doppler Ultrasonic Diagnostic Instruments (Vet) Lito key Press it to activate the lito function. Biopsy key Press it to activate the biopsy function. Menu key Press it to hide/display the menu function. Body mark key Press it to enter/exit the body mark status.
Page 15
User’s Manual V1.01 Full Digital Color Doppler Ultrasonic Diagnostic Instruments (Vet) B mode key Press it to enter the B mode. Color mode key Press it to enter/exit the color flow Doppler mode. PDI mode key Press it to enter the PDI mode. PW mode key Press it to enter/exit the pulsed wave Doppler mode.
User’s Manual V1.01 Full Digital Color Doppler Ultrasonic Diagnostic Instruments (Vet) ● Rotate it to adjust the calibrated angle in the PW mode. ● Rotate it to delete or restore the last trace during the trace measurement. Used for increasing or decreasing the depth of the Depth key image.
User’s Manual V1.01 Full Digital Color Doppler Ultrasonic Diagnostic Instruments (Vet) 1. LOGO 2. Owner name 3. Hospital Name 4. Direction mark of the probe 5. Gender 6. Patient ID and patient's type 7. Probe model 8. MI and TIS indices 9.
User’s Manual V1.01 Full Digital Color Doppler Ultrasonic Diagnostic Instruments (Vet) Chapter 3 System Configuration 3.1 Typical configuration 1. Main unit 1 unit 2. 6.5 MHz micro-convex probe 1 PC 3. 7.5 MHz high frequency linear array probe 1 PC 4.
Warning: 1. When instrument works abnormally, do stop working, turn off the power and check the reason, then contacts the KAIXIN Company about it. 2. Turn off power and pull out of the plug from socket after each ultrasonic diagnostic operation.
User’s Manual V1.01 Full Digital Color Doppler Ultrasonic Diagnostic Instruments (Vet) Warning: The probes and foot switch as described in this section may be replaced by operator. To replace according to the method of this section, does not result in an unacceptable risk.
User’s Manual V1.01 Full Digital Color Doppler Ultrasonic Diagnostic Instruments (Vet) Attention: 1. Turn off the ultrasonic system before disconnecting the probe. Disconnecting the probe with system power on may damage the system or probe. 2. Before disconnecting the probe, place the probe on the probe holder and put its cable on the probe cable hanger so that the probe may not be damaged or injury person by unexpected fall.
User’s Manual V1.01 Full Digital Color Doppler Ultrasonic Diagnostic Instruments (Vet) 5.4 Connect to power There are two ways to power on the instrument: mians supply and battery. The device's battery is automatically charged while the power is connected to the mains supply. Once the charge is completed, the power supply will no longer continue to charge the battery.
User’s Manual V1.01 Full Digital Color Doppler Ultrasonic Diagnostic Instruments (Vet) 2. Connect the printer to the mains supply with power cable, with AV video cable connect composite video output interface ( ) of the system to the printer video input interface. You can use the video printer when the connection is completed.
User’s Manual V1.01 Full Digital Color Doppler Ultrasonic Diagnostic Instruments (Vet) 5.8 Ultrasonic probe check before and after operation Check if there are any exceptionally on the surface of the probe or cable jacket, such as peeling, cracks, bulge, or if the acoustic lens is reliable, disinfected or cleaned. 5.9 Main unit check before and after operation 5.9.1 Inspection before start-up Check the following items before starting the machine:...
User’s Manual V1.01 Full Digital Color Doppler Ultrasonic Diagnostic Instruments (Vet) Chapter 6 Presetting System The Preset menu allows you to specify general system settings, printer, probe, scan, measurement method, text annotation and DICOM server. All your customized settings can be saved and can run even after you rebooting the system.
Page 27
User’s Manual V1.01 Full Digital Color Doppler Ultrasonic Diagnostic Instruments (Vet) Figure 6-2 System Preset Screen Item Descriptions Hospital Name Enter the hospital name. Language Choose a system language of the user interface. Set the date format of the system. Date Format Options: YYYY/MM/DD, MM/DD/YYYY or DD/MM/YYYY.
User’s Manual V1.01 Full Digital Color Doppler Ultrasonic Diagnostic Instruments (Vet) Set the animal types. Options: Cat, Dog, Cattle, Horse, Sheep, Other. Type Set the animal type need to restart the machine, the animal type can take effect. Set the functions of shortcut key. P1~ P2 Options: Save Film, Colorize Set the system date and time.
User’s Manual V1.01 Full Digital Color Doppler Ultrasonic Diagnostic Instruments (Vet) Steps for editing the exam type are as follows: 1. Click the drop-down menu beside [Probe name] to select the desired probe. 2. Select the exam type in the list of [The exam of system] and click > to add the selected exam type to the list of [The exam of probe].
User’s Manual V1.01 Full Digital Color Doppler Ultrasonic Diagnostic Instruments (Vet) Figure 6-5 Measurement Preset Screen Item Descriptions Set the heart rate. Options: 1-8. Enable or disable the dotted line between two markers. Cursor Line Display 6.5 Presetting the Annotations You can create, edit or delete the defined text annotations in the Comment Preset screen.
User’s Manual V1.01 Full Digital Color Doppler Ultrasonic Diagnostic Instruments (Vet) Figure 6-6 Annotation Preset Screen To add an annotation Click the [Exam Mode] to select one exam type. Input the characters beside [Add] and click [Add] to add the defined annotation to the list of [The comment of custom].
Page 32
User’s Manual V1.01 Full Digital Color Doppler Ultrasonic Diagnostic Instruments (Vet) Figure 6-7 DICOM Preset Screen Item Descriptions System title The system title is the same as the local AE title by default. Organ name Enter the hospital's name. AE title Enter the name of AE title and the system title changes with it.
User’s Manual V1.01 Full Digital Color Doppler Ultrasonic Diagnostic Instruments (Vet) table above. 3. Click [Validate] to verify the connection between the system and the sever. A prompt box will pop up to indicate the verification information. 4. If the connection is successful, you can save the patient information and images to the server. Attention: Ensure that the system is successfully connected to the DICOM server before use.
User’s Manual V1.01 Full Digital Color Doppler Ultrasonic Diagnostic Instruments (Vet) Chapter 7 Patient Information You should enter the relevant information about the patient, select the appropriate probe and exam type before examining. Complete patient information helps to define the exam. Attention: System time and the current time inconsistency may cause misdiagnosis.
User’s Manual V1.01 Full Digital Color Doppler Ultrasonic Diagnostic Instruments (Vet) Figure 7-2 Probe and Exam Type Selection Screen Attention: Probe types and icons display on the LED screen when the probe are connected. If the LED screen displays nothing, check the connections of the probe. If problem still exists, please stop using the ultrasound system immediately and contact the manufacturer or the authorized service engineer.
User’s Manual V1.01 Full Digital Color Doppler Ultrasonic Diagnostic Instruments (Vet) Chapter 8 Optimizing the Image After the patient information, probe and exam type are selected, you can select the imaging mode to optimize the image for exact diagnosis. 8.1 B Mode B mode is the most frequently used two-dimension imaging mode.
Page 37
User’s Manual V1.01 Full Digital Color Doppler Ultrasonic Diagnostic Instruments (Vet) Method 2: Select the B mode parameter on the menu at the top left of the screen by using the Select knob, press Select knob change state between “parameter item select state” and “parameter item adjust state”, rotate the Select knob change parameter item for “parameter item select state”...
Page 38
User’s Manual V1.01 Full Digital Color Doppler Ultrasonic Diagnostic Instruments (Vet) Steps: Move cursor to [Freq] menu item and press Set key to adjust the frequency. Rotate Select knob to highlight the [Freq] menu item, press Select knob item to confirmation select [Freq] item, and then rotate Select knob to adjust the frequency of the probe.
Page 39
User’s Manual V1.01 Full Digital Color Doppler Ultrasonic Diagnostic Instruments (Vet) Move cursor to [Scan Density] menu item and press Set key to adjust it. Rotate Select knob to highlight the [Scan Density] menu item, press Select knob item to confirmation select [Scan Density] item, and then rotate Select knob to adjust it.
User’s Manual V1.01 Full Digital Color Doppler Ultrasonic Diagnostic Instruments (Vet) Rotate Select knob to highlight the [Image Enhance] menu item, press Select knob item to confirmation select [Image Enhance] item, and then rotate Select knob to adjust it. 8.1.2.17 Adjusting the Acoustic Power Acoustic power is adjusted to increase or decrease the transmitting frequency of the probe.
User’s Manual V1.01 Full Digital Color Doppler Ultrasonic Diagnostic Instruments (Vet) Figure 8-2 Color Doppler Mode Image 3. Adjust the color ROI Move the trackball to adjust the position of the color ROI. Press the Set key and move the trackball to adjust the size of the ROI, or press the Set key again to adjust the position of ROI.
Page 42
User’s Manual V1.01 Full Digital Color Doppler Ultrasonic Diagnostic Instruments (Vet) Steps: Move cursor to [Scale] menu item and press Set key to adjust it. Rotate Select knob to highlight the [Scale] menu item, press Select knob item to confirmation select [Scale] item, and then rotate Select knob to adjust it.
Page 43
User’s Manual V1.01 Full Digital Color Doppler Ultrasonic Diagnostic Instruments (Vet) 8.2.2.10 Adjusting the Scan Density Density improves the resolution and decreases frame rate. Therefore, you have to balance the frame rate and the image quality when adjusting the line density. Steps: ...
User’s Manual V1.01 Full Digital Color Doppler Ultrasonic Diagnostic Instruments (Vet) 8.3 PDI Mode PDI (Power Doppler Imaging) is a color flow imaging technology which adds the flow signal in the Color mode image. PDI uses the number and amplitude of red blood cell going through in the flow to create the color-coded imaging.
User’s Manual V1.01 Full Digital Color Doppler Ultrasonic Diagnostic Instruments (Vet) 8.4 M Mode M mode is generally used for cardiac applications. In the M mode, you can position the M-line in the 2D image on the anatomy of interest, and then learn about the tissue motion along that line in an M-mode trace.
Page 46
User’s Manual V1.01 Full Digital Color Doppler Ultrasonic Diagnostic Instruments (Vet) Parameters Adjustment in the M Mode 8.4.2.1 Adjusting the Gain M gain controls the overall brightness of the M trace. Steps: Rotate the Gain knob clockwise to increase the value. ...
Page 47
User’s Manual V1.01 Full Digital Color Doppler Ultrasonic Diagnostic Instruments (Vet) Rotate Select knob to highlight the [Gray Map] menu item, press Select knob item to confirmation select [Gray Map] item, and then rotate Select knob to adjust it. 8.4.2.7 Adjusting the Chroma Chroma is used to colorize the grayscale image.
User’s Manual V1.01 Full Digital Color Doppler Ultrasonic Diagnostic Instruments (Vet) 8.5 PW Mode Pulsed Wave Doppler (PW) is a Doppler mode that measures velocity in a PW sample volume and displays that information in a spectral trace with audio output. 8.5.1 Entering the PW Mode Steps for entering the PW mode are as follows: 1.
User’s Manual V1.01 Full Digital Color Doppler Ultrasonic Diagnostic Instruments (Vet) 5. Optimize the PW mode image. For details, refer to Section 8.5.2 Optimizing the PW Mode Image. Press the Save IP key to save the parameter settings in the PW mode as the default settings. Press key on the keyboard to adjust the Doppler sound.
Page 50
User’s Manual V1.01 Full Digital Color Doppler Ultrasonic Diagnostic Instruments (Vet) 8.5.2.4 Adjusting the Baseline Baseline represents the position of zero velocity or frequency. The velocity range in one direction can be increased or decreased by adjusting the baseline. Steps: ...
Page 51
User’s Manual V1.01 Full Digital Color Doppler Ultrasonic Diagnostic Instruments (Vet) Steps: Move cursor to [Steer] menu item and press Set key to adjust the angle of the spectral Doppler line. Rotate Select knob to highlight the [Steer] menu item, press Select knob item to confirmation select [Steer] item, and then rotate Select knob to adjust the angle of the spectral Doppler line.
User’s Manual V1.01 Full Digital Color Doppler Ultrasonic Diagnostic Instruments (Vet) Chapter 9 Processing the Images You can work with the acquired images by using the features provided by the ultrasound system, such as reviewing, magnifying and freezing the image, cropping the cine, and making annotations on the image. 9.1 Displaying the image 9.1.1 Dual-splitting Display You can position two images side by side on the screen to compare the images acquired at different times.
User’s Manual V1.01 Full Digital Color Doppler Ultrasonic Diagnostic Instruments (Vet) 3. Repeatedly press to freeze the real time scan and unfreeze other images. The probe direction mark is highlighted in real time scan and becomes gray in the frozen mode. 4.
User’s Manual V1.01 Full Digital Color Doppler Ultrasonic Diagnostic Instruments (Vet) You can review or crop the cine. 9.4.1 Reviewing the Cine Reviewing it manually You can review the cine frame by frame by using the trackball. Reviewing it automatically Press [Auto Play] menu item to playback the cine automatically and you can adjust the review speed during playback.
User’s Manual V1.01 Full Digital Color Doppler Ultrasonic Diagnostic Instruments (Vet) 1. Press the Body Mark key on the control panel to activate the body mark annotation and a body mark appears on the screen. 2. Select one body mark by using the trackball and press the Set key to confirm. The selected body mark displays on the bottom left of the screen.You can also adjust the direction of the probe by rotating the Angle knob.
User’s Manual V1.01 Full Digital Color Doppler Ultrasonic Diagnostic Instruments (Vet) Chapter 10 Measurements and Calculations You can perform measurements on the ultrasound image, and the system automatically calculates the measurement results. You can use the results to conclude the precise diagnostic information. General measurements and specific measurements are provided.
User’s Manual V1.01 Full Digital Color Doppler Ultrasonic Diagnostic Instruments (Vet) 10.2 Measurement Menu In the real-time scanning or freezing state, press Caliper to pop up the general measurement menu, press Measure key to pop up the specific measurement menu. Taking small parts measurement as an example to introduce the specific measurement menu: Exam type Measurement...
User’s Manual V1.01 Full Digital Color Doppler Ultrasonic Diagnostic Instruments (Vet) 10.5 General Measurement The measurement results of the general measurements are not recorded in the measurement report, but all the specific measurements are based on the general measurements. General measurements can be performed in the B mode, M mode, Color/PDI mode and PW mode. The operations of Color/PDI mode and the B mode measurements are similar.
Page 59
User’s Manual V1.01 Full Digital Color Doppler Ultrasonic Diagnostic Instruments (Vet) of the screen. 2. Select [Circle/Area] on the menu by moving the trackball and press Set to confirm. Select[Trace] on the pop-up menu. A marker displays on the image. 3.
Page 60
User’s Manual V1.01 Full Digital Color Doppler Ultrasonic Diagnostic Instruments (Vet) 2. Select [Volume] on the menu by moving the trackball and press Set to confirm. Select[Ellipse] on the pop-up menu. A marker displays on the image. 3. Move the marker to the starting point by using the trackball, press Set to fix the marker and the second marker displays.
User’s Manual V1.01 Full Digital Color Doppler Ultrasonic Diagnostic Instruments (Vet) 2. Select [Ratio(D)] on the menu by moving the trackball and press Set to confirm. A marker displays on the image. 3. Move the marker to the starting point by using the trackball, press Set to fix the marker and the second marker displays.
User’s Manual V1.01 Full Digital Color Doppler Ultrasonic Diagnostic Instruments (Vet) 1. Press the M key on the control panel to enter the M mode. 2. Press the Caliper key to activate the measurement, and the measurement menu displays on the top left of the screen.
User’s Manual V1.01 Full Digital Color Doppler Ultrasonic Diagnostic Instruments (Vet) 10.5.4 PW Mode Measurements Flow speed ratio, speed, time, heart rate, acceleration and spectrum trace measurements can be performed in the PW mode. You can also press [BC Caliper] to enter the general measurements in the B mode. For details, please refer to Section 10.5.1 B Mode Measurements.
Page 64
User’s Manual V1.01 Full Digital Color Doppler Ultrasonic Diagnostic Instruments (Vet) 3. Select [HR] on the menu by moving the trackball and press Set to confirm. A marker displays on the image. 4. Move the marker to the starting point by using the trackball, press Set to fix the marker and the second marker displays.
User’s Manual V1.01 Full Digital Color Doppler Ultrasonic Diagnostic Instruments (Vet) 4. Move the marker to the starting point of peak waveform of the end systole by using the trackball, press Set to fix the marker and the second marker displays. 5.
User’s Manual V1.01 Full Digital Color Doppler Ultrasonic Diagnostic Instruments (Vet) RK W LK L LK D Left kidney Refer to Section 10.5.1.1 Distance. LK W Kidney measurement Steps: 1. Press the Measure key in the real time or frozen B/Color/PDI mode, and the measurement menu displays on the top left of the screen.
User’s Manual V1.01 Full Digital Color Doppler Ultrasonic Diagnostic Instruments (Vet) 10.6.2.2 GA/EDD The gestational age (GA) and expected date of deliver (EDD) can be calculated by measuring BD, HD, CRL, GSD, TD, HSD-H, GSD-V and BPD. 10.6.2.3 Obstetrics Measurements in the PW Mode Steps: 1.
Page 68
User’s Manual V1.01 Full Digital Color Doppler Ultrasonic Diagnostic Instruments (Vet) Left Ventricle Diastolic Refer to Section 10.5.1.1 LVLd Major Axis Distance. LV long axis area at Refer to Section 10.5.1.2 Area LVLd end-diastole and Circumference. Left Ventricle/ Single Plane Method Left Ventricle Systolic Refer to Section 10.5.1.1 LVLs...
Page 69
User’s Manual V1.01 Full Digital Color Doppler Ultrasonic Diagnostic Instruments (Vet) Calculation Item Description Formula End- diastolic Left Ventricular EDV(ml)=(8/3/ )× LVAMd( )× LVALd( )/LVIDd(cm) Volume End- systolic Left Ventricular ESV(ml)=( 8/3/ ) × LVAMs( )× LVALs( )/LVIDs(cm) Volume Stroke Volume SV = EDV(ml) - ESV(ml) Ejection Fraction EF=100%×...
Page 70
User’s Manual V1.01 Full Digital Color Doppler Ultrasonic Diagnostic Instruments (Vet) Mod.Simpson This measurement method calculates the LV volume by using the LV long axis view, the LV short axis at the level of the mitral value and the LV short axis view at level of the papillary muscles. Steps: 1.
Page 71
User’s Manual V1.01 Full Digital Color Doppler Ultrasonic Diagnostic Instruments (Vet) the image. 3. Perform the LVPWdIVSTd LVIDd measurements one by one and the measurement results displays in the result box. Measurement Measurement Description Measurement Method Item Sub-item LVPWd LV posterior wall thickness at end diastole Refer to section10.5.1.1 Interventricular septal thickness at end Left Ventricle...
Page 72
User’s Manual V1.01 Full Digital Color Doppler Ultrasonic Diagnostic Instruments (Vet) 10.6.3.7 FS FS can be calculated by measuring LVIDd and LVIDs. Steps: 1. Press the Measure key in the real time or frozen B/Color/PDI mode, and the measurement menu displays on the top left of the screen.
Page 73
User’s Manual V1.01 Full Digital Color Doppler Ultrasonic Diagnostic Instruments (Vet) End-diastolic Left Ventricular EDV(ml)=(7× LVIDd )/(2.4 + LVIDd(cm)) Volume Fractional Shortening FS = 100%× (LVIDd(cm)- LVIDs(cm))/ LVIDd(cm) End-systolic Left Ventricular ESV(ml)=(7× LVIDs )/(2.4+LVIDs(cm)) Volume Stroke Volume SV = EDV(ml) - ESV(ml) Ejection Fraction EF=100%×...
Page 74
User’s Manual V1.01 Full Digital Color Doppler Ultrasonic Diagnostic Instruments (Vet) 10.6.3.10 Mitral Valve Steps: 1. Press the Measure key in the real time or frozen M mode, and the measurement menu displays on the top left of the screen. 2.
User’s Manual V1.01 Full Digital Color Doppler Ultrasonic Diagnostic Instruments (Vet) 3. A marker displays on the image. Perform the measurement according to the following table and the measurement results displays in the result box. Measurement Item Measurement Method Refer to Section 10.5.4.6 Spectrum Trace. Upper Ext Vein Lower Ext Vein 10.6.5 Small Parts Measurements...
User’s Manual V1.01 Full Digital Color Doppler Ultrasonic Diagnostic Instruments (Vet) Chapter 11 Report The report will be generated after you perform the specific measurements but not the general measurements. It contains patient information, measurement results, findings and prompt. 11.1 Reviewing the Report Press the Report key on the control panel to enter the Report screen.
User’s Manual V1.01 Full Digital Color Doppler Ultrasonic Diagnostic Instruments (Vet) Chapter 12 Managing Images/Data You can save images after scanning the images. The saved patient data can be backup to the USB storage devices for future use. 12.1 Saving the Image/Cine Attention: The data is saved to the built-in disk of the system by default.
User’s Manual V1.01 Full Digital Color Doppler Ultrasonic Diagnostic Instruments (Vet) Saved patient 1. Press the Station key on the control panel to enter the Station screen. Figure 12-2 Working Station 2. Select one patient in the [Exam List] by using the trackball and press the Set key, the thumbnail images display in the [File List].
User’s Manual V1.01 Full Digital Color Doppler Ultrasonic Diagnostic Instruments (Vet) 12.5.1 Saving to USB Storage Devices Steps are as follows: 1. Connect the USB storage device to the system and the bottom right of the screen displays the icon of USB.
User’s Manual V1.01 Full Digital Color Doppler Ultrasonic Diagnostic Instruments (Vet) Chapter 13 Puncture guide 13. 1 Operation of puncture guide line Attention: During the operation of puncture guide, do not freeze the image. 1. In the real time B mode, press the Biopsy key on the control panel, the Biopsy menu is displayed at the top left of the screen.
2. Clean the probe with sterile water to remove the residual chemicals after disinfection, because the residual chemicals may be harmful for humans. 3. Kaixin Company will not make any guarantee for the efficacy of disinfector. Please contact the appropriate manufacturer for details.
Page 83
User’s Manual V1.01 Full Digital Color Doppler Ultrasonic Diagnostic Instruments (Vet) Glutaraldehyde-based disinfectant: Chemical Name Reagent Name Step Cidex Glutaraldehyde Please refer to the instructions Glutaraldehyde (2.4%) disinfectant of the solution for details. Non-glutaraldehyde-based disinfectant: Chemical Name Reagent Name Step Phthalaldehyde solution Please refer to the instructions Cidex OPA...
User’s Manual V1.01 Full Digital Color Doppler Ultrasonic Diagnostic Instruments (Vet) 14.1.2 Clean the probe cable and its connector 1. Clean the probe cable and its connector with soft, dry cloth. 2. In case of die-hard blots, clean with soft cloth dipped in moderate detergent and then air-dry it. 14.1.3 Clean the LED screen Use a soft cloth dipped in glass cleaner to clean the LED screen, and then air-dried.
User’s Manual V1.01 Full Digital Color Doppler Ultrasonic Diagnostic Instruments (Vet) Attention: Do not throw the battery into water or be wet, which will lead to the battery leakage, explosion or fire; Do not use or store the battery near the heat source, such as fire or heater, which will lead to the battery leakage, explosion or fire;...
If any abnormity occur, timely contact International Trade Dept of Kaixin for support. If the following problems occur on starting up the machine, try to make corrections following the method in the table. If the problem remains unsolved, contact International Trade Dept of Kaixin for support. Trouble...
14.7 Maintenance by service personnel To ensure the performance and safety of instrument, it must be checked after using 1 year. When check the instrument, please consult the International Trade Dept of Kaixin or its dealers, as they need to have professional technology engineers.
User’s Manual V1.01 Full Digital Color Doppler Ultrasonic Diagnostic Instruments (Vet) Chapter 15 Storage and Transportation Storage and Transportation 1. If the instrument is stored over 3 months, take out the instrument from the packing case, connect it to power supply for 4 hours, and then disconnect the power and place it in the case again following the direction indicated by arrows on the package.
Page 89
KAIXIN ELECTRONIC XUZHOU KAIXIN ELECTRONIC INSTRUMENT CO., LTD. Kaixin Mansion, C-01, Economic Development Zone, Xuzhou, Jiangsu, China Post Code: 221004 Tel: +86-516-87732932/87733758 Fax: +86-516-87732932/87792848 Website: http://www.kxele.com E-mail: info@kxele.com Information contained in this manual is subject to change without further notice.