Kaixin DCU12 User Manual

Kaixin DCU12 User Manual

Full digital color doppler ultrasonic diagnostic instrument (vet)
Table of Contents

Advertisement

KAIXIN
DCU12
User's Manual
Full Digital Color Doppler Ultrasonic
Diagnostic Instrument (Vet)
XUZHOU KAIXIN Electronic Instrument Co., Ltd.

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the DCU12 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Kaixin DCU12

  • Page 1 KAIXIN Full Digital Color Doppler Ultrasonic DCU12 Diagnostic Instrument (Vet) User’s Manual XUZHOU KAIXIN Electronic Instrument Co., Ltd.
  • Page 2 Statement Kaixin has the final explanation right of this user’s manual. Kaixin was considered responsible for the safety, reliability and performance in case of meeting all the following requirements: 1. Assembly, expansion, readjustment, improve and repair are all performed by professionals recognized by Kaixin;...
  • Page 3 This user’s manual contains warnings for predictable dangers. Users shall also exercise care at any time to be aware of the dangers unforeseen in this manual. Kaixin shall not be liable for the damages and losses arising out of neglecting to follow the operation instructions herein described.
  • Page 4 Full Digital Color Doppler Ultrasonic Diagnostic Instrument (Vet) User’s Manual V1.00 Safety Cautions 1. Warning Symbols and Definitions The following warning symbols are used in this manual to indicate safety level and other important items. Please remember these symbols and understand the meaning as you read this user’s manual. These symbols convey specific meanings as detailed in the table below: Symbols &...
  • Page 5 Full Digital Color Doppler Ultrasonic Diagnostic Instrument (Vet) User’s Manual V1.00 3. Labels Label Description Danger: 1. It may have explosion hazard if used with flammable gas. 2. Operate the probe with care. Read the probe information in the relevant manual for proper probe use.
  • Page 6: Table Of Contents

    Full Digital Color Doppler Ultrasonic Diagnostic Instrument (Vet) User’s Manual V1.00 Contents Chapter One Technical Specifications……………………………………………………………1 Chapter Two System Outline………………………………………………………………………3 2.1 Structure composition of the instrument ………………………………………………3 2.2 Components name…………………………………………………………………………3 2.3 Parts of the probe…………………………………………………………………….. 3 2.4 Rear panel instruction…………………………………………………………………….4 2.5 Function keys instruction…………………………………………………………………4 2.6 Trackball……………………………………………………………………………………..…6 Chapter Three...
  • Page 7 Full Digital Color Doppler Ultrasonic Diagnostic Instrument (Vet) User’s Manual V1.00 9.1 Select probe and application mode…………………………………………………………..18 9.2 User Settings………………………………………………………………………………..18 Chapter Ten Image operation modes…………………………………………………………19 10.1 Operation modes……………………………………………………………………………..19 10.2 Switch operation modes……………………………………………………………………..19 10.3 Image parameter adjustment……………………………………………………………..19 10.3.1 Image brightness adjustment……………………………………………………………..19 10.3.2 B/2B mode………………………………………………………………………………..20 10.3.3 B/M mode………………………………………………………………………………..21 10.3.4 M mode…………………………………………………………………………………..23...
  • Page 8 Full Digital Color Doppler Ultrasonic Diagnostic Instrument (Vet) User’s Manual V1.00 19.3 Formats of image files……………………………………………………………………..64 19.4 Review, demo image/video/report………………………………………………………..64 19.5 Dicom archive……………………………………………………………………………..64 Chapter Twenty Edit and print report…………………………………………………………65 20.1 Enter/exit report editing module…………………………………………………………..65 20.2 Work interface of report editing…………………………………………………………….. 65 Chapter Twenty-one Print and Transfer files…………………………………………………67 21.1 Print…………………………………………………………………………………………..67...
  • Page 9: Chapter One Technical Specifications

    Full Digital Color Doppler Ultrasonic Diagnostic Instrument (Vet) User’s Manual V1.00 Chapter One Technical Specifications 1.1 Technical parameters 1. Gray scale: 256 2. Monitor: 10.4” LED 3. Adapter rating: 100-240V~, 1.2-0.6A, 50-60Hz 4. Output of Adapter: DC12.8V 3.0A 5. Main device rating: DC12V 3.0A 6.
  • Page 10 Full Digital Color Doppler Ultrasonic Diagnostic Instrument (Vet) User’s Manual V1.00 35. Cine playback function; 36. Body marks function; 37. Information annotation function; 38. User defined function; 39. Input patient information function; 40. Automatically general report function; 41. Recall report by one-key; 42.
  • Page 11: Chapter Two System Outline

    Full Digital Color Doppler Ultrasonic Diagnostic Instrument (Vet) User’s Manual V1.00 Chapter Two System Outline 2.1 Structure composition of the instrument DCU12 full digital color Doppler ultrasonic diagnostic instrument (Vet) is composed of main unit and probe and etc. 2.2 Components name Probe Display...
  • Page 12: Rear Panel Instruction

    Full Digital Color Doppler Ultrasonic Diagnostic Instrument (Vet) User’s Manual V1.00 2.4 Rear panel instruction USB port USB port Color SVGA output Network interface Color SVGA Out Video output interface Foot switch interface Fig. Real panel A sketch map Power input interface Fig.
  • Page 13 Full Digital Color Doppler Ultrasonic Diagnostic Instrument (Vet) User’s Manual V1.00 Value adjustment Rotate this knob to adjust the corresponding parameters; Menu-item Rotate this knob to select the menu-item; selection Press the key and then rotate【Value】knob, adjust the scanning Depth depth;...
  • Page 14: Trackball

    Full Digital Color Doppler Ultrasonic Diagnostic Instrument (Vet) User’s Manual V1.00 Cine playback In frozen status, press the key to enter the cine playback status; Measurement and Press the key to enter general measurement menu and calculation application measurement menu; ·In B/B mode, realize the switching of dual B mode;...
  • Page 15: Chapter Three System Configuration

    Full Digital Color Doppler Ultrasonic Diagnostic Instrument (Vet) User’s Manual V1.00 Chapter Three System Configuration 3.1 Typical configuration Main unit 1 unit Power adapter 1 PC 3.5MHz convex array probe 1 PC 6.5MHz micro convex array probe 7.5 MHz high frequency linear array probe Choose one 6.5MHz animal transrectal linear array probe 3.2 Optional parts...
  • Page 16: Chapter Five System Installation And Check

    Warning: 1. When instrument works abnormally, do stop working, turn off the power and check the reason, then contacts the KAIXIN Company about it. 2. Turn off power and pull out of the plug from socket after each ultrasonic diagnostic operation.
  • Page 17: System Placement

    Full Digital Color Doppler Ultrasonic Diagnostic Instrument (Vet) User’s Manual V1.00 5.1 System placement Please carefully read through and fully understand the safety cautions before moving and placing the system. 1. Unpack the instrument case and check the goods for its completeness according to the packing list furnished.
  • Page 18: Install Foot Switch

    SUPPLY MAINS. Not to position the EQUIPMENT so that it is difficult to operate the disconnection device. 2. The device should be used only with power adapter provided by Kaixin Company. 3. To avoid damaging power adapter or harming people by unexpected fallen, make sure the power adapter is placed on the leveled plane.
  • Page 19: System Reset

    Full Digital Color Doppler Ultrasonic Diagnostic Instrument (Vet) User’s Manual V1.00 5.8.2 Inspection after start-up Check the following items after starting the machine: 1. No abnormal voice, strange smell and overheating appear. 2. Check the machine to ensure a normal start-up: The power indication light is on and startup picture is shown on the screen.
  • Page 20: Chapter Six Work Main Interface

    Full Digital Color Doppler Ultrasonic Diagnostic Instrument (Vet) User’s Manual V1.00 Chapter Six Work Main Interface 6.1 Startup and Shutdown First connect the power supply, and then press the button on the bottom of screen to start the machine, it will display the B-mode image on the screen. If you want to turn off the machine, please press the button again, then pull out mains plug from the supply mains.
  • Page 21: Menu Adjustment Region

    Full Digital Color Doppler Ultrasonic Diagnostic Instrument (Vet) User’s Manual V1.00 “Probe selection” key. 8. Current application mode Display the current using application mode, such as: abdomen. 6.4 Menu adjustment region Mainly display control menus of image adjustment and optimization parameters, and through relevant keys and trackball to adjust and control;...
  • Page 22: Chapter Seven System Preset

    Full Digital Color Doppler Ultrasonic Diagnostic Instrument (Vet) User’s Manual V1.00 Chapter Seven System Preset System preset function is used to set the system running environment, status and configuration parameters in each of application mode. The parameters of system preset are stored within the system memory, them will not be lost when shutdown and power-off.
  • Page 23: Probe Information

    Full Digital Color Doppler Ultrasonic Diagnostic Instrument (Vet) User’s Manual V1.00 1. Move the cursor to the “Waiting Time” input box, press『Set』key to confirm the cursor position; 2. Input the screensaver time, number (0-99), in minute. Note: Go beyond the system setting energy-saving time without pressing any key, the system will automatically enter the energy saving status.
  • Page 24: Store Preset

    Warning: Business personnel or service personnel setting the password, be sure to remember the set password, make notes! If you forget password, consult Kaixin Company. 7.10 About , press 『Set』 key, the screen will pop up the “About” dialog box. In the dialog 『...
  • Page 25: Chapter Eight Patient Information Input

    Full Digital Color Doppler Ultrasonic Diagnostic Instrument (Vet) User’s Manual V1.00 Chapter Eight Patient Information Input Under normal conditions, user checks the patient without needing to input patient information. This system has built-in patient information management. To achieve the management, query and storage functions for the data information of patient, recommend user to input patient information as complete as possible before examination.
  • Page 26: Chapter Nine Select Probe And Application Mode

    Full Digital Color Doppler Ultrasonic Diagnostic Instrument (Vet) User’s Manual V1.00 Chapter Nine Select probe and application mode Attention: When the markers of measurement and calculation, text comments are being used, if you switch the application mode, the markers of measured data, text comments and arrows on the image will be cleared, measured data will not be saved.
  • Page 27: Chapter Ten Image Operation Modes

    Full Digital Color Doppler Ultrasonic Diagnostic Instrument (Vet) User’s Manual V1.00 Chapter Ten Image operation modes Warning: The images provided in this system only for reference when doctor diagnosis, it is not in charge of the accuracy of diagnostic results. The inspection doctor is responsible for the accuracy of examination results.
  • Page 28: B/2B Mode

    Full Digital Color Doppler Ultrasonic Diagnostic Instrument (Vet) User’s Manual V1.00 10.3.2 B/2B mode In B mode, in the “Parameters” menu user can adjust “Dyn Range”, “ Edge”, “Comp Curve”, “IP”, “Frame Avg”, “Image Opt”; in the “Display” menu user can adjust “V Reverse”, “H Reverse”, “ESWL Line”, “Quad”.
  • Page 29: B/M Mode

    Full Digital Color Doppler Ultrasonic Diagnostic Instrument (Vet) User’s Manual V1.00 Select the different image effects; the Image 12-level, 0~ default is 4 when connecting 3.5MHz optimization probe. V Reverse On/Off “ ” is direction marker. Under normal circumstances, the marker “ ” of image H Reverse On/Off direction is in the upper-left corner of...
  • Page 30 Full Digital Color Doppler Ultrasonic Diagnostic Instrument (Vet) User’s Manual V1.00 Edge enhancement is to enhance the Edge image outline, so to strengthen the image 4-level, 0~3 enhancement edge display, the value is greater, and the enhancement is more obvious. System built-in some gray-scale curves, to adapt to the needs of different checking parts and application range.
  • Page 31: M Mode

    Full Digital Color Doppler Ultrasonic Diagnostic Instrument (Vet) User’s Manual V1.00 Switch work frequency Harmonics 2-level harmonic frequency. Acoustic 4-level AP25, AP50, AP75, AP100 Power 10.3.4 M mode Display adjustable parameters in M mode For M images, through “Parameters” menu user can adjust “Dyn Range”, “ Edge”, “Comp Curve”, “IP”, “M speed”...
  • Page 32 Full Digital Color Doppler Ultrasonic Diagnostic Instrument (Vet) User’s Manual V1.00 the smaller sampling depth is. The display area of two-dimensional image and color flow sampling ranges also has an impact on the frame rate. If the real-time in the examination is poor, you can improve frame rate by the following methods: A.
  • Page 33 Full Digital Color Doppler Ultrasonic Diagnostic Instrument (Vet) User’s Manual V1.00 Color Afterglow increases, low-velocity and 4-level, 0~3 Persistence low-flow blood flow is easily displayed. Color Limits the display of color scale, eliminates 6-level, 0~5 Threshold a lot of unnecessary noise. Color Balance 16-level, 0~15 Adjust color brightness.
  • Page 34: Pdi Mode

    Full Digital Color Doppler Ultrasonic Diagnostic Instrument (Vet) User’s Manual V1.00 6.5L64H1 probe variable frequency range:4.5/5.5/6.5/7.5 7.5L40H1 probe variable frequency range:6.0/6.5/7.5/9.0 By default, color focus position Focus Focus position automatically locates in the vertical center of color sampling box. ·Operate in CFM, PDI and PW mode; ·The function only be used with linear array Linear color probe;...
  • Page 35 Full Digital Color Doppler Ultrasonic Diagnostic Instrument (Vet) User’s Manual V1.00 First, sampling detection should be on the major axis of the vessel, not on the minor axis or inclined axis, so the ratio between the detected vessel length and diameter is as large as possible. Second, the major axis of the vessel should be parallel with the scanning section, the angle between beam and blood flow is as small as possible, the maximum not more than 60 degrees.
  • Page 36 Full Digital Color Doppler Ultrasonic Diagnostic Instrument (Vet) User’s Manual V1.00 flow signals that low echo intensity and high velocity can pass. Reduce noise; increase the sensitivity of Sensitivity 6-level, 0~5 blood flow. Color Afterglow increases, low-velocity and 4-level, 0~3 Persistence low-flow blood flow is easily displayed.
  • Page 37 Full Digital Color Doppler Ultrasonic Diagnostic Instrument (Vet) User’s Manual V1.00 In B/PW mode, turn【Function】knob to adjust the gain of CFM image; press 【Update】key to enter PW mode, turn Gain 0~127dB 【Function】knob to adjust the gain of PW image. According to the different probes, the levels 3.5MHz: of depth are different.
  • Page 38: Chapter Eleven Cine Playback

    Full Digital Color Doppler Ultrasonic Diagnostic Instrument (Vet) User’s Manual V1.00 Chapter Eleven Cine playback In frozen status, system automatically enters the manual playback status of cine mode, user toggles the trackball to view the images frame by frame, can measure, text annotate, body mark and other operations for the valuable images;...
  • Page 39: Chapter Twelve Synopsis Of Measurement Menu

    Full Digital Color Doppler Ultrasonic Diagnostic Instrument (Vet) User’s Manual V1.00 Chapter Twelve Synopsis of Measurement Menu Attention: 1. During the measurement process, if switch the exam mode or probe, all the measurements and marks in the image will be cleared. 2.
  • Page 40: Measurement Menu Composition

    Full Digital Color Doppler Ultrasonic Diagnostic Instrument (Vet) User’s Manual V1.00 PW measurement menu: Can get PW measurement report by measuring the velocity, heart rate, time, acceleration, Pulsatility Index (PI)/Resistive Index (RI), maximum pressure, mean pressure, ratio, CO, flow, etc. Attention: The accuracy of software measurement: distance measurement ≤...
  • Page 41 Full Digital Color Doppler Ultrasonic Diagnostic Instrument (Vet) User’s Manual V1.00 Move the trackball, press『Set』key, locate the measurement cursor to the measured starting point, move the cursor, pull-out measurement line, while measured results immediately refresh and show in the region of measured results. Now, if press key, you can switch the starting point and end point of the measurement.
  • Page 42: Chapter Thirteen B Mode Measurement And Calculation

    Full Digital Color Doppler Ultrasonic Diagnostic Instrument (Vet) User’s Manual V1.00 Chapter Thirteen B mode measurement and calculation 13.1 B general measurements B general measurement is usually done in B image mode. B measurement menu as figures: Fig. B measurement menu 13.1.1 Distance measurement Function: measures the distance between two points.
  • Page 43 Full Digital Color Doppler Ultrasonic Diagnostic Instrument (Vet) User’s Manual V1.00 B. Ellipse method measurement Function: measures the circumference/area of a closed region with the ellipse approximations. 1. Move the cursor to〖Ellipse〗submenu of〖Cir/Area〗in【B Meas】menu, press『Set』key; 2. Operate the trackball or the direction keys to move the “+” mark to the desired position, press『Set』 key and make the “+”...
  • Page 44 Full Digital Color Doppler Ultrasonic Diagnostic Instrument (Vet) User’s Manual V1.00 to make it satisfy the area need to be measured, press『Measure』key to exchange the starting point and end point; press 『 Set』 key, the measured value is automatically displayed at the built-in characters “Gmax: ”...
  • Page 45: B Abdominal Measurements

    Full Digital Color Doppler Ultrasonic Diagnostic Instrument (Vet) User’s Manual V1.00 13.1.9 Depth measurement 1. In B mode choose the valuable image; 2. Move the cursor to〖Depth〗menu item in【B Meas】menu, press『Set』key, the cursor shows “+”; 3. Operate the trackball or the direction keys to move the “+” mark to the end point of the measurement, the “Depth1: mm”...
  • Page 46 Full Digital Color Doppler Ultrasonic Diagnostic Instrument (Vet) User’s Manual V1.00 Gallbladder-Width GB-W B mode distance measurement Gallbladder-Thickness GB-T B mode distance measurement cystic duct diameter Choledocho B mode distance measurement Extrahepatic Bile ducts Ext.hepatic Bile B mode distance measurement diameter Intrahepatic Bile duct Int.hepatic Bile...
  • Page 47: B Obstetric Measurements

    Full Digital Color Doppler Ultrasonic Diagnostic Instrument (Vet) User’s Manual V1.00 2. The detailed operation refers to the “Distance” measurement of B general measurement; 3. The corresponding measurement results will be shown on right side of the display area; 4. Repeat steps one to three to measure for next time. 13.2.6 Abdominal report After finishing the measurement and calculation of abdominal examination, system will generate the abdominal examine diagnosis report.
  • Page 48 Full Digital Color Doppler Ultrasonic Diagnostic Instrument (Vet) User’s Manual V1.00 In B/M, CFM or PDI mode and in frozen status, press key to quickly open【B Ob Meas】menu; or press key, choose【B Ob Meas】item from “B Meas” menu. 13.3.1 Measurement and calculation items It can calculates gestational weeks (GW) and estimated date of childbirth (EDC) for bovine, canine, equine, feline, goat, llama, ovine and swine.
  • Page 49: B Cardiac Measurements

    Full Digital Color Doppler Ultrasonic Diagnostic Instrument (Vet) User’s Manual V1.00 3. Measure fat thickness and loin, method is the same as the “Distance” measurement of “B Meas”; 4. The result of %Lean will be showed on the right of the shown area; 5.
  • Page 50 Full Digital Color Doppler Ultrasonic Diagnostic Instrument (Vet) User’s Manual V1.00 Mitral valve horizontal left ventricular short B mode “Circumference/area” ellipse LVAMd axis area, at end diastole [mm or cm measurement Left Ventricular internal diameter, at end LVIDd B mode distance measurement diastole [mm or cm] Left Ventricular long axis area, at end B mode “Circumference/area”...
  • Page 51 Full Digital Color Doppler Ultrasonic Diagnostic Instrument (Vet) User’s Manual V1.00 Left Ventricular long axis diameter, at end LVLd B mode distance measurement diastole [mm or cm] Papillary muscles horizontal left ventricular B mode “Circumference/area” ellipse LVAPd short axis area, at end diastole [mm2 or cm2] measurement Mitral valve horizontal left ventricular short B mode “Circumference/area”...
  • Page 52 Full Digital Color Doppler Ultrasonic Diagnostic Instrument (Vet) User’s Manual V1.00 13.4.2 Single plane ellipse measurement LV function Measurement method: 1. Move the cursor to 〖LV〗 menu item, pop up its submenu. Move the cursor to 〖S-P Ellipse〗 submenu, press『Set』key , the cursor will show “+”. 2.
  • Page 53: B Urology Measurements

    Full Digital Color Doppler Ultrasonic Diagnostic Instrument (Vet) User’s Manual V1.00 Key in height and weight: 1. Move the cursor to 〖 Key In〗 menu item, pop up its submenu. Move the cursor to 〖 H & W〗 submenu, press『Set』key, pop up “Input Height & Weight” dialog box in the center of screen; 2.
  • Page 54 Full Digital Color Doppler Ultrasonic Diagnostic Instrument (Vet) User’s Manual V1.00 13.5.1 Items of measurement and calculation of urology examination Name Abbreviation Algorithm and Operation Unit Prostate Volume Prostate-V (PV) PV=0.52×[PV-L]×[PV-W]×[PV-T]/1000 PV-Length PV-L B mode distance measurement PV-Width PV-W B mode distance measurement PV-Thickness PV-T B mode distance measurement...
  • Page 55: B Superficial Organs Measurements

    Full Digital Color Doppler Ultrasonic Diagnostic Instrument (Vet) User’s Manual V1.00 13.5.6 Kidney Volume (Kidney-V) 1. Choose〖Kidney-V〗in【B Uro Meas】menu, press『Set』key to enter measurement status. 2. The whole measurement includes Kidney-L, Kidney-W and Kidney-T. The detailed operation of every measurement refers to the “Distance” measurement of B general measurement. 3.
  • Page 56 Full Digital Color Doppler Ultrasonic Diagnostic Instrument (Vet) User’s Manual V1.00 Fig. Superficial organs measurement menu 13.6.1 Items of measurement and calculation of superficial organs examination Name Abbreviation Algorithm and Operation Unit Lt Lobe-V = 0.523×[L.Lobe-L] ×[L.Lobe-W] Left Lobe Volume Lt Lobe-V ×[L.Lobe-T]/1000 Lt Lobe V-Length...
  • Page 57 Full Digital Color Doppler Ultrasonic Diagnostic Instrument (Vet) User’s Manual V1.00 1. Choose〖Isthmus Thk〗in【B Sup Meas】menu, press『Set』key to enter measurement status; 2. The detailed operation refers to the “Distance” measurement of B general measurement; 3. After finishing measurement, the isthmus thickness will be showed at the right of display area. 13.6.5 Left Gland Thickness (Lt Gland) Measurement method: 1.
  • Page 58: Chapter Fourteen M Mode Measurement And Calculation

    Full Digital Color Doppler Ultrasonic Diagnostic Instrument (Vet) User’s Manual V1.00 Chapter Fourteen M mode measurement and calculation 14.1 M general measurements In B/M, M mode, general measurement can measure depth, slope, heart rate, time and ratio. M general measurement menu as figures: Fig.
  • Page 59: M Cardiac Measurements

    Full Digital Color Doppler Ultrasonic Diagnostic Instrument (Vet) User’s Manual V1.00 4. Press『Set』key to finish the measurement; 5. Repeat steps two to four to measure next time. 14.1.4 Time measurement Function: Measure the time interval between two points. Measurement method: 1.
  • Page 60 Full Digital Color Doppler Ultrasonic Diagnostic Instrument (Vet) User’s Manual V1.00 End-systolic left ventricular volume ESV[mL]=7*LVIDs [cm] /(2.4+LVIDs [cm]) [mL] Stroke Volume [mL] SV[mL]= EDV[mL]- ESV[mL] Cardiac Output [L/min] CO[L/min]= SV[mL]* HR[bpm]/1000 Ejection Fraction EF[No unit]= SV[mL]/ EDV[mL] Fractional Shortening FS=(LVIDd[mm]-LVIDs[mm])/ LVIDd[mm] Stroke Volume Index SI= SV[mL]/ BSA[m...
  • Page 61 Full Digital Color Doppler Ultrasonic Diagnostic Instrument (Vet) User’s Manual V1.00 Mitral valve volume QMV[mL]=4*DEV[cm/s]*DCT[s] Left Atrium inside diameter M mode distance measurement Aorta inside diameter M mode distance measurement LAD/AOD Inside diameter ratio of left atrium to aorta LAD/AOD[no unit]= LAD [mm]/AOD [mm] MAVO1 Aorta valve begin opening diameter M mode distance measurement...
  • Page 62 Full Digital Color Doppler Ultrasonic Diagnostic Instrument (Vet) User’s Manual V1.00 Attention: 1. It should key in or measure the essential parameters of heart rate, ET, height, weight and others before measuring the cardiac function parameters, or else it will show incomplete data.
  • Page 63: Chapter Fifteen Cfm/Pdi Mode Measurement And Calculation

    Full Digital Color Doppler Ultrasonic Diagnostic Instrument (Vet) User’s Manual V1.00 1. Move the cursor to〖CO〗menu item, press『Set』key to enter measurement status; 2. Measure time, heart rate with “Distance” measurement method; 3. After finishing the measurement, CO will be shown on the right side of the display area. 14.2.6 ET measurement After finishing LV function measurement, it can calculate MVCF by measuring left ventricle ET.
  • Page 64: Chapter Sixteen Pw Mode Measurement And Calculation

    Full Digital Color Doppler Ultrasonic Diagnostic Instrument (Vet) User’s Manual V1.00 Chapter Sixteen PW mode measurement and calculation 16.1 PW mode measurements Press key, and then press【Update】key to enter PW spectrum mode, freeze the image, press key, it can measure the following measurements, as figures: Fig.
  • Page 65 Full Digital Color Doppler Ultrasonic Diagnostic Instrument (Vet) User’s Manual V1.00 minimum velocity in a cycle, measure the velocities of two points and calculate Vmean, SD ratio, PI and RI. Measure method: 1. Move the cursor to 〖 PI 〗 menu item in 【 PW Meas 】 menu, press 『 Set 』 key to enter measurement status; 2.
  • Page 66 Full Digital Color Doppler Ultrasonic Diagnostic Instrument (Vet) User’s Manual V1.00 2. Operate the trackball or the direction keys to move the cursor to the point at the maximum velocity to be measured, press 『 Set 』 key to fix the point; 3.
  • Page 67: Pw Measurement Report

    Full Digital Color Doppler Ultrasonic Diagnostic Instrument (Vet) User’s Manual V1.00 3. Move the cursor to 〖 Flow 〗 submenu of 〖 Flow 〗 in 【 PW Meas 】 menu, press 『 Set 』 key, at last finish “Flow”, the right display area shows the Flow measured value. This measurement ends. 16.2 PW measurement report After finishing the PW measurement and calculation, system will automatically generate PW examine report.
  • Page 68: Chapter Seventeen Text Comment And Arrow Mark

    Full Digital Color Doppler Ultrasonic Diagnostic Instrument (Vet) User’s Manual V1.00 Chapter Seventeen Text comment and arrow mark In the diagnosis process, usually require to add comments or markers on the ultrasonic image, to increase the readability of ultrasonic images. Comments can be added on the frozen image. Attention: Must ensure the correctness of comments.
  • Page 69: Delete Text Comment

    Full Digital Color Doppler Ultrasonic Diagnostic Instrument (Vet) User’s Manual V1.00 17.2.2 Auto comment entries After system enters into text comment status, you can select the desired annotation entries from the auto annotation entries showed in the left side, and input it on the image. Operation method: 1.
  • Page 70: Chapter Eighteen Body Mark, Puncture Guide And Lithotripsy

    Full Digital Color Doppler Ultrasonic Diagnostic Instrument (Vet) User’s Manual V1.00 Chapter Eighteen Body Mark, Puncture guide and Lithotripsy Body mark indicates the probing part and probe scanning direction when patient receiving inspection, it plays an illustration role for the image. Attention: Must ensure the correctness of body mark.
  • Page 71: Chapter Nineteen Image Management

    Full Digital Color Doppler Ultrasonic Diagnostic Instrument (Vet) User’s Manual V1.00 Chapter Nineteen Image management This system provides image management function; user can view, delete or export the stored images/video within the system, and this function helps user to do offline diagnosis, offline presentations and other demands for obtained ultrasonic images.
  • Page 72: Formats Of Image Files

    Full Digital Color Doppler Ultrasonic Diagnostic Instrument (Vet) User’s Manual V1.00 Export: Images, reports or videos in the folder of D disk are saved to the folder with same name in the U disk; Print: Print the selected image or report; the printing operation method refers to the Chapter twenty-one “Print and Transfer files”...
  • Page 73: Chapter Twenty Edit And Print Report

    Full Digital Color Doppler Ultrasonic Diagnostic Instrument (Vet) User’s Manual V1.00 Chapter Twenty Edit and print report The system built-in ultrasound work station report function, assist user in finishing ultrasonic examination, and according to a certain format to output standard ultrasonic graphic report. 20.1 Enter/exit report editing module In the main interface of ultrasonic work, press freeze key, press key to enter ultrasound report...
  • Page 74 Full Digital Color Doppler Ultrasonic Diagnostic Instrument (Vet) User’s Manual V1.00 also can add stored pictures from D disk or U disk if need; (2) Add pictures from D disk or U disk. Move the trackball, select the desired pictures from the file list at the left side, press 『...
  • Page 75: Chapter Twenty-One Print And Transfer Files

    Direct Line to the Ultrasound Schematic: Direct Line to the Ultrasound 1. Open the DCU12 Full Digital Color Doppler Ultrasonic Diagnostic Instrument. 2. Use a network cable to connect the network port of the instrument and the computer's network port, and then installs the printer on your computer.
  • Page 76 Full Digital Color Doppler Ultrasonic Diagnostic Instrument (Vet) User’s Manual V1.00 key on the keyboard of the instrument to open 【 System Preset 】 menu, move the cursor to 5. Press , press 『 Set 』 key to show the “Net Preset” dialog box, the default setting as shown on 『...
  • Page 77 Line to the network Schematic: Line to the network 1. Open the DCU12 Full Digital Color Doppler Ultrasonic Diagnostic Instrument. 2. Using a network cable to connect the network port of the instrument to the switch, the network port of computer also connects to the switch, and then installs the printer on your computer, the printer sets for sharing;...
  • Page 78 Full Digital Color Doppler Ultrasonic Diagnostic Instrument (Vet) User’s Manual V1.00 Figure 1. Network Setting on the way “Line to the network” Note: In the interface “Ultrasound network param setting”, if not display the local IP, please wait for a while to display. key on the keyboard of the instrument to open 【...
  • Page 79 Full Digital Color Doppler Ultrasonic Diagnostic Instrument (Vet) User’s Manual V1.00 Figure 3. Network Setting on “Line to the network” 7. After the setting is complete, move the cursor to select “Verify Settings” button marked on the LAN Server Set, press 『 Set 』 key, it will pop up success prompt box, press 「 OK 」 in the prompt box, the LAN server set is successful, the icon at the lower right corner of the screen will changed from , it indicates that the network is successful, you can achieve the network printing function.
  • Page 80: Usb Print

    Full Digital Color Doppler Ultrasonic Diagnostic Instrument (Vet) User’s Manual V1.00 21.1.2 USB Print 1. Insert U disk into the USB port at the back of main unit; 2. When U disk connected to this instrument, the status bar at the bottom right of the screen will pop up USB icon;...
  • Page 81: Chapter Twenty-Two System Maintenance

    2. Clean the probe with sterile water to remove the residual chemicals after disinfection, because the residual chemicals may be harmful for humans. 3. Kaixin Company will not make any guarantee for the efficacy of disinfector. Please contact the appropriate manufacturer for details.
  • Page 82: Clean The Probe Cable And Its Connector

    Full Digital Color Doppler Ultrasonic Diagnostic Instrument (Vet) User’s Manual V1.00 Glutaraldehyde-based disinfectant: Chemical Name Reagent Name Step Cidex Glutaraldehyde Please refer to the instructions of Glutaraldehyde (2.4%) disinfectant the solution for details. Non-glutaraldehyde-based disinfectant: Chemical Name Reagent Name Step Phthalaldehyde solution Please refer to the instructions of Cidex OPA...
  • Page 83: Clean The Trackball

    Full Digital Color Doppler Ultrasonic Diagnostic Instrument (Vet) User’s Manual V1.00 Attention: Do not clean the screen with hydrocarbon detergent such as alcohol or OA equipment cleaning media. These kinds of liquid may degrade the internal function of the screen. Attention: Prohibit using sharp objects to touch the LED screen, and prohibit pressing or squeezing against the LED screen.
  • Page 84: Data Maintenance

    22.5 Maintenance by service personnel To ensure the performance and safety of instrument, it must be checked after using 1 year. When check the instrument, please consult the International Trade Dept of Kaixin or its dealers, as they need to have professional technology engineers.
  • Page 85: Chapter Twenty-Three Storage And Transportation

    Full Digital Color Doppler Ultrasonic Diagnostic Instrument (Vet) User’s Manual V1.00 Chapter Twenty-three Storage and Transportation Storage and Transportation 1. If the instrument is stored over 3 months, take out the instrument from the packing case, connect it to power supply for 4 hours, and then disconnect the power and place it in the case again following the direction indicated by arrows on the package.
  • Page 86: Appendix B System Block Diagram

    Full Digital Color Doppler Ultrasonic Diagnostic Instrument (Vet) User’s Manual V1.00 Appendix A System Block Diagram Probe socket High-voltage switch Probe ID identify driver circuit High-voltage pulse generating circuit Pulse generation and driver circuit High-voltage isolation System Adapter AC IN power A/D converter module...
  • Page 87 KAIXIN ELECTRONIC XUZHOU KAIXIN ELECTRONIC INSTRUMENT CO., LTD Kaixin Mansion, C-01. Economic Development Zone, Xuzhou, Jiangsu, China Post Code: 221004 Tel: +86-516-87732932/87733758 Fax: +86-516-87732932/87792848 Website: http://www.kxele.com E-mail: gm01@kxele.com Information contained in this manual is subject to change without further notice.

Table of Contents