Okolab LEO 2.0 Technical Notes

Gas measurement in heracell incubators

Advertisement

Quick Links

RECOMMENDED
DEVICE SETUP
IN LEO 2.0
FOR GAS
MEASUREMENT
PRECAUTIONS:
• Please be cautious with your cultured embryos, considering that during
the measurement procedure the composition of the gas in the chamber
may be affected.
• Read the manuals for the incubator and LEO 2.0 for usage and safety
details.
STEP 1
Connect TUBE-A to LEO 2.0's
HERACELL
HERACELL
TUBE-A
TUBE-B
IN
port and to the Moisure Trap, and TUBE-B from Moisture Trap to the
gas Sample out port.
gas Sample out port
MOISTURE
TRAP
TECHNICAL NOTE
LEO 2.0 gas measurement
in
HERACELL
N2
Evaluate the need for chamber cleaning and disinfection
after working on the incubator.
As good practice, it is recommended to have LEO 2.0
charged before use, and not to charge it during use.
Always read the manual of the devices.
If LEO 2.0 shows a message that it's warming up, it is
advised to complete the warm up period before using.
For Time Logging Measurement remove the Trap Top Lid.
1
2
incubators
CO2
%
O2
%
3

Advertisement

loading
Need help?

Need help?

Do you have a question about the LEO 2.0 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Okolab LEO 2.0

  • Page 1 Always read the manual of the devices. may be affected. • Read the manuals for the incubator and LEO 2.0 for usage and safety If LEO 2.0 shows a message that it’s warming up, it is advised to complete the warm up period before using.
  • Page 2 10 minutes, proceed to the Next step. STEP 4 Start a LEO 2.0 gas measurement. S U PP O RT@ O KO -L AB .COM O KO -L A B .CO M TN-LEO 2.0-01rev01-EN_How to measure gas in HERACELL...