P P r r o o C C u u i i t t y y B B e e d d S S e e r r i i e e s s O O p p e e r r a a t t i i o o n n s s M M a a n n u u a a l l 300900000000 3009PX-L-100 3009PX-L-200...
Page 3
S S y y m m b b o o l l s s Refer to instruction manual/booklet Consult instructions for use General warning Caution Warning; electricity Fuse rating Non-ionizing radiation Warning; crushing of hands China RoHS with declarable substances Catalogue number Serial number European medical device...
Page 4
Authorized representative in the European Community Authorized representative in Switzerland For US Patents see www.stryker.com/patents Manufacturer Date of manufacture Safe working load Mass of equipment NAWI Class IIII Maximum patient weight Adult patient Alternating current Direct current Duty cycle of product ≤2m...
Page 5
T T a a b b l l e e o o f f C C o o n n t t e e n n t t s s Warning/Caution/Note Definition ..........................3 Summary of safety precautions ..........................3 Introduction ................................6 Product description ..............................6 Indications for use..............................6 Clinical benefits ..............................7...
Page 6
Pendant, basic option............................42 Pendant, advanced option .............................43 Footboard control panel - Home ..........................45 Footboard control panel - Position ..........................45 Footboard control panel, advanced - Home option ....................46 Footboard control panel - Motion lock ........................48 Footboard control panel, advanced - Motion lock option ...................49 Footboard control panel - Scale..........................50 Setting the scale...............................50 Weighing a patient............................51...
Page 7
This product is equipped with a hospital-grade plug for protection against electric shock hazard. • Always use a Stryker supplied interface cable. Use of any other cable may cause the product to not function as intended, which may result in patient or user injury.
Page 8
• Do not apply the brakes to slow or stop the product while the product is in motion. • Always unplug the power cord before you transport the product. • Always release the brakes before you transport the product. Do not transport the product with the brakes applied. •...
Page 9
• Do not pinch the power cord in the bed frame. • Do not use the siderails as a push or pull device. Always move the product using the integrated handles in the headboard and footboard. • Always remove the patient helper before you transport the product. •...
Page 10
P P r r o o d d u u c c t t d d e e s s c c r r i i p p t t i i o o n n The Stryker P P r r o o C C u u i i t t y y ™ bed series is a powered, adjustable hospital bed used in combination with a patient support surface.
Page 11
S S p p e e c c i i f f i i c c a a t t i i o o n n s s W W A A R R N N I I N N G G - - Always use Stryker approved support surfaces that have been tested for compatibility with the product frame to avoid the risk of patient entrapment.
Page 12
12 VDC, 1.2 Ah (x2) (Stryker part number: 700000341245) N N o o t t e e - - Always replace with Stryker approved batteries. Z Z o o o o m m battery voltage 12 VDC, 20 Ah (x2) (Stryker part number: 700000341246) N N o o t t e e - - Always replace with Stryker approved batteries.
Page 13
8 - 10.5 in. 20.3 - 26.7 cm Stryker reserves the right to change specifications without notice. Specifications listed are approximate and may vary slightly from product to product or by power supply fluctuations. O O p p e e r r a a t t i i o o n n...
Page 14
S S u u b b s s t t a a n n c c e e o o f f v v e e r r y y h h i i g g h h c c o o n n c c e e r r n n D D e e s s c c r r i i p p t t i i o o n n N N u u m m b b e e r r ( ( S S V V H H C C ) ) c c h h e e m m i i c c a a l l n n a a m m e e...
Page 15
2.4 GHz: All channels supported Channel plan 5 GHz: All channels supported (Recommend against the use of DFS and ISM channels) Leverage hospital SSID Support for 802.11r Other Support for Cisco CCX (Fast roaming) S S p p e e c c i i f f i i c c a a t t i i o o n n - - C C h h i i p p s s e e t t Q Q C C A A 9 9 3 3 7 7 7 7 - - 3 3 ( ( Q Q u u a a l l c c o o m m m m A A t t h h e e r r o o s s ) ) I I t t e e m m U U n n i i t t B B a a n n d d...
Page 16
The client uses an additional 5-25 KB per device for each subscription that is created by Stryker (S S E E M M /i i B B e e d d V V i i s s i i o...
Page 17
Add session timeout of at least 24 Recommended hours Microwave ovens Avoid using microwave ovens near Recommended Stryker wireless products as this will cause degradation of service or no service due to electromagnetic or radio-frequency (RF) interference ( Wireless coexistence notifications (page 76)).
Page 18
C C o o n n t t a a c c t t i i n n f f o o r r m m a a t t i i o o n n Contact Stryker Customer Service or Technical Support at: 1-800-327-0770.
Page 19
To view your operations or maintenance manual online, see https://techweb.stryker.com/. Have the serial number (A) of your Stryker product available when calling Stryker Customer Service or Technical Support. Include the serial number in all written communication. S S e e r r i i a a l l n n u u m m b b e e r r l l o o c c a a t t i i o o n n You can find the serial number (A) below the headboard at the head of the bed (Figure 3).
Page 20
W W A A R R N N I I N N G G • Always use a Stryker supplied interface cable. Use of any other cable may cause the product to not function as intended, which may result in patient or user injury.
Page 21
B B e e d d Locator or S S e e c c u u r r e e C C o o n n n n e e c c t t , see the i i B B e e d d Locator Installation Manual or S S e e c c u u r r e e C C o o n n n n e e c c t t Installation/Configuration Manual. For any set up questions, contact Stryker Technical Support at: •...
Page 22
• If the product is equipped with the patient helper option, the product must be in a flat position and either be 17.5 in. (44.5 cm) from the wall at a 14 in. (35.6 cm) product height (or more) or within 3 ft (0.9 m) from the wall at a 22 in. (55.9 cm) product height (or more) to establish a first connection.
Page 23
O O p p e e r r a a t t i i o o n n P P l l u u g g g g i i n n g g o o r r u u n n p p l l u u g g g g i i n n g g t t h h e e p p r r o o d d u u c c t t W W A A R R N N I I N N G G •...
Page 24
Always check battery backup function. Replace the battery if it does not perform as intended during preventive maintenance. To charge the battery: 1. Plug the power cord into the wall outlet. 2. Turn on the battery switch (A) at the head end of the product (Figure 6). F F i i g g u u r r e e 6 6 –...
Page 25
T T r r a a n n s s p p o o r r t t i i n n g g t t h h e e p p r r o o d d u u c c t t W W A A R R N N I I N N G G •...
Page 26
F F i i g g u u r r e e 8 8 – – A A p p p p l l y y i i n n g g t t h h e e b b r r a a k k e e s s F F i i g g u u r r e e 9 9 –...
Page 27
F F i i g g u u r r e e 1 1 0 0 – – A A p p p p l l y y i i n n g g S S t t e e e e r r - - L L o o c c k k F F i i g g u u r r e e 1 1 1 1 –...
Page 28
N N o o t t e e • The throttle indicator will flash green when it senses your hand on the handle. The throttle indicator is solid green when Z Z o o o o m m is ready. •...
F F i i g g u u r r e e 1 1 3 3 – – Z Z o o o o m m D D r r i i v v e e R R e e l l e e a a s s e e S S t t o o p p p p i i n n g g t t h h e e p p r r o o d d u u c c t t w w i i t t h h Z Z o o o o m m d d r r i i v v e e o o p p t t i i o o n n You can rotate the Z Z o o o o m m throttle to stop the product in three ways.
To lower the Z Z o o o o m m handles, press the handle release (F) on the base of the handle and lower the handle (Figure 15). Z Z o o o o m m d d r r i i v v e e m m o o d d e e s s N N o o t t e e - - Push the product in neutral.
Page 31
F F i i g g u u r r e e 1 1 5 5 – – Z Z o o o o m m t t h h r r o o t t t t l l e e Throttle Rotates to move the bed Green (flashing)
Page 32
A A c c t t i i v v a a t t i i n n g g t t h h e e C C P P R R r r e e l l e e a a s s e e h h a a n n d d l l e e C C A A U U T T I I O O N N - - Always confirm that all persons and equipment are away from the area below and around the Fowler before you activate the CPR release handle.
Page 33
2. Lower the footboard until the footboard seats into the sockets. R R a a i i s s i i n n g g t t h h e e s s i i d d e e r r a a i i l l s s W W A A R R N N I I N N G G •...
Page 34
F F i i g g u u r r e e 1 1 8 8 – – S S i i d d e e r r a a i i l l i i n n t t e e r r m m e e d d i i a a t t e e p p o o s s i i t t i i o o n n L L o o w w e e r r i i n n g g t t h h e e s s i i d d e e r r a a i i l l s s W W A A R R N N I I N N G G •...
Page 35
F F i i g g u u r r e e 1 1 9 9 – – S S i i d d e e r r a a i i l l l l o o w w e e s s t t p p o o s s i i t t i i o o n n E E x x t t e e n n d d i i n n g g o o r r r r e e t t r r a a c c t t i i n n g g t t h h e e b b e e d d e e x x t t e e n n d d e e r r The bed extender allows you to extend the length of the product by 12 in.
Page 36
F F i i g g u u r r e e 2 2 0 0 – – R R e e s s t t r r a a i i n n t t s s t t r r a a p p t t i i e e - - i i n n l l o o c c a a t t i i o o n n s s S S e e c c u u r r i i n n g g t t h h e e r r o o l l l l b b e e l l t t s s t t r r a a p p s s o o p p t t i i o o n n The roll belt brackets are located on each side of the product below the siderails.
Page 37
S S e e c c u u r r i i n n g g a a F F o o l l e e y y b b a a g g t t o o t t h h e e F F o o l l e e y y b b a a g g h h o o o o k k There are two Foley bag hooks under the foot section (A) (Figure 22), one on either side of the product.
Page 38
I I s s o o l l i i b b r r i i u u m m P P E E s s u u p p p p o o r r t t s s u u r r f f a a c c e e o o p p t t i i o o n n The I I s s o o l l i i b b r r i i u u m m PE is a powered support surface with features that provide pressure redistribution, low air loss, turn assist, max inflate, and lateral rotation.
Page 39
Returns to the H H o o m m e e screen ( Footboard control panel, advanced - Home Home option (page 46)) Displays menu functions ( Footboard control panel, advanced - Menu Menu option (page 61)) Select to increase the redistribution firmness setting Firm Default redistribution firmness setting Medium...
Page 40
• Soft (E) ( Footboard control panel, advanced - Pressure option (page 34)) • Medium (D) ( Footboard control panel, advanced - Pressure option (page 34)) Firm (C) ( Footboard control panel, advanced - Pressure option (page 34)) • A A c c t t i i v v a a t t i i n n g g m m a a x x i i n n f f l l a a t t e e o o p p t t i i o o n n To activate max inflate: 1.
Turn assist, patient left Selects turn assist for the patient left side Select 15 seconds, 30 minutes, or 120 minutes for the turn assist hold Hold time time Start Starts selected turn assist hold time Locks or unlocks selected mattress operation settings Lock Displays current mattress status Status bar...
Page 42
Returns to the H H o o m m e e screen ( Footboard control panel, advanced - Home Home option (page 46)) Displays menu functions ( Footboard control panel, advanced - Menu Menu option (page 61)) Full rotation. Available when Fowler angle is -5° to 35° Full Reduced rotation.
Page 43
• Do not zero or tare the scale or weigh the patient during operation. • Always center the patient on the support surface and check frequently to maintain proper position. N N o o t t e e • A healthcare professional should determine the use of lateral rotation outside of the weight range of 60 lb (27.2 kg) to 460 lb (208.6 kg).
Bed Exit Press and hold to cancel Bed Exit alarm Nurse call Activates nurse call Gatch up Raises the Gatch Gatch down Lowers the Gatch Fowler up Raises the Fowler Fowler down Lowers the Fowler Fowler 30°+ Illuminates when the Fowler is 30°+ Bed height up Raises the litter Bed height down...
Page 45
Bed Exit Press and hold to cancel Bed Exit alarm Nurse call Activates nurse call Places the product into a position for patient ingress Patient stand assist or egress Cardiac chair position Places the product into the cardiac chair position Vascular position Places the product into the vascular position Places the product into the Trendelenburg position...
Page 46
• Always route the cables, wires, and tubing from other equipment so that parts of the product do not pinch them. C C A A U U T T I I O O N N - - Always clear the product of obstacles before you use motion functions. Healthcare professionals must instruct patients how to operate the patient control panel.
Page 47
Nurse call Activates nurse call Fowler up Raises the Fowler Fowler down Lowers the Fowler Gatch up Raises the Gatch Gatch down Lowers the Gatch P P e e n n d d a a n n t t , , a a d d v v a a n n c c e e d d o o p p t t i i o o n n W W A A R R N N I I N N G G •...
Page 48
Nurse call Activates nurse call Fowler up Raises the Fowler Fowler down Lowers the Fowler Gatch up Raises the Gatch Gatch down Lowers the Gatch Volume up Increases the volume Volume down Decreases the volume Channel up Changes the channel up Changes the channel down Channel down Mute...
Page 49
F F o o o o t t b b o o a a r r d d c c o o n n t t r r o o l l p p a a n n e e l l - - H H o o m m e e The home functions are located on the touch panel of the footboard.
The P P o o s s i i t t i i o o n n screen displays the position functions of the product. Returns to the H H o o m m e e screen (see Footboard control Home panel - Home (page 45)) Low height range...
Page 51
Displays pressure functions option ( Footboard control panel, advanced - Pressure Pressure option (page 34)) Displays turn functions option ( Footboard control panel, advanced - Turn Turn option (page 36)) Displays pulmonary functions option ( Footboard control panel, advanced Pulmonary - Pulmonary option (page 37)) Basic - Displays brake lock status Advanced option - Apply or release the brakes ( Applying or releasing the...
Places the product into a 0° horizontal position Bed flat Fowler up Raises the Fowler Fowler down Lowers the Fowler Gatch up Raises the Gatch Gatch down Lowers the Gatch Bed height up Raises the litter Bed height down Lowers the litter Places the product into the Trendelenburg position (head down with foot Trendelenburg Places the product into the Reverse Trendelenburg position (head up...
Lock all Locks or unlocks all motion function Back Returns to the M M o o t t i i o o n n L L o o c c k k screen N N o o t t e e •...
Page 54
N N o o t t e e • The CPR release handle overrides all lockouts. • If the product is in a specific position when you enable a lock, the product is locked in that position. • Lock parameters are saved when you unplug the product. •...
Page 55
2. On the S S c c a a l l e e screen, press the Z Z e e r r o o or T T a a r r e e button (D) ( Footboard control panel - Scale (page 50)). 3.
Page 56
Returns to the S S c c a a l l e e screen ( Footboard control panel Return - Scale (page 50)) Select to reset the added equipment to zero Reset Select to save the current displayed weight as Save equipment To add or remove equipment:...
Page 57
Zero (Non-NAWI) Select to zero or tare the scale or set the scale for a new Tare (NAWI) patient ( Setting the scale, advanced option (page 53)) Exit Returns to the S S c c a a l l e e screen S S e e t t t t i i n n g g t t h h e e s s c c a a l l e e , , a a d d v v a a n n c c e e d d o o p p t t i i o o n n Before you place a patient on the product, zero or tare the scale.
Page 58
Returns to the H H o o m m e e screen ( Footboard control panel, Home advanced - Home option (page 46)) Returns to the S S c c a a l l e e screen ( Footboard control panel, Return advanced - Scale option (page 52)) Select to reset the added equipment to zero...
Page 59
Returns to the H H o o m m e e screen ( Footboard control Home panel - Home (page 45)) Allows the patient to move freely on the bed. Alarms when the patient moves 50 percent body weight out of the designated zone.
Page 60
To disarm Bed Exit, press T T u u r r n n O O f f f f (E) ( Footboard control panel - Bed Exit (page 54)). F F o o o o t t b b o o a a r r d d c c o o n n t t r r o o l l p p a a n n e e l l , , a a d d v v a a n n c c e e d d - - B B e e d d E E x x i i t t o o p p t t i i o o n n The B B e e d d E E x x i i t t screen displays the Bed Exit functions of the product.
A A r r m m i i n n g g o o r r d d i i s s a a r r m m i i n n g g B B e e d d E E x x i i t t , , a a d d v v a a n n c c e e d d o o p p t t i i o o n n W W A A R R N N I I N N G G - - Do not use Bed Exit to replace patient monitoring protocol.
Page 62
Monitors head left siderail in the highest position Head left siderail Foot right siderail Monitors foot right siderail in the highest position Monitors foot left siderail in the highest position Foot left siderail Head Angle Monitors head angle of the bed Monitors bed flat position Bed Flat Turn Off...
Page 63
Returns to the H H o o m m e e screen ( Footboard control Home panel, advanced - Home option (page 46)) Low Height Monitors low height of the bed Foot right siderail Monitors foot right siderail in the highest position Head right siderail Monitors head right siderail in the highest position Monitors head left siderail in the highest position...
Page 64
• LEDs on the footboard and siderails flash red • Sound alarm is triggered To disable i i B B e e d d Watch, press T T u u r r n n O O f f f f (I) ( Footboard control panel, advanced - i i B B e e d d Watch option (page 58)). F F o o o o t t b b o o a a r r d d c c o o n n t t r r o o l l p p a a n n e e l l - - M M e e n n u u The M M e e n n u u screen displays the functions available to view or change for the product.
Page 65
S S e e t t t t i i n n g g s s Scale info Displays the current software version, local gravity, and calibrated gravity Wi-Fi info Displays the MAC address, connection type, IP address, SSID, signal strength, and BSSID i i B B e e d d Locator Displays the locator ID and locator battery status...
Displays lockout functions ( Footboard control panel, advanced - Motion Motion lock lock option (page 49)) Select to reset the bed configuration Soft Reset History Displays the S S c c a a l l e e H H i i s s t t o o r r y y screen Activates and displays Bed Exit functions ( Footboard control panel, Bed Exit advanced - Bed Exit option (page 56))
Page 67
F F i i g g u u r r e e 2 2 3 3 – – L L o o w w h h e e i i g g h h t t r r a a n n g g e e A A d d v v a a n n c c e e d d S S e e t t t t i i n n g g s s i i B B e e d d Watch preset Select to choose the i i B B e e d d Watch presets...
Page 68
N N o o t t e e - - The screen will disappear in 30 seconds if you do not press O O K K . 3. You can return to normal product motion. 3009-009-001 Rev AG.0...
Page 69
A A c c c c e e s s s s o o r r i i e e s s a a n n d d p p a a r r t t s s These accessories and parts may be available for use with your product. Confirm availability for your configuration or region.
Page 70
F F i i g g u u r r e e 2 2 5 5 – – H H A A V V A A S S U U t t w w o o - - s s t t a a g g e e I I V V p p o o l l e e A A t t t t a a c c h h i i n n g g o o r r r r e e m m o o v v i i n n g g t t h h e e p p a a t t i i e e n n t t h h e e l l p p e e r r o o p p t t i i o o n n W W A A R R N N I I N N G G - - Always use two people to attach or remove the patient helper.
Page 71
F F i i g g u u r r e e 2 2 6 6 – – A A t t t t a a c c h h i i n n g g o o r r r r e e m m o o v v i i n n g g t t h h e e p p a a t t i i e e n n t t h h e e l l p p e e r r Reverse steps to remove the patient helper.
Page 72
F F i i g g u u r r e e 2 2 7 7 – – A A d d j j u u s s t t i i n n g g t t h h e e p p a a t t i i e e n n t t h h e e l l p p e e r r A A t t t t a a c c h h i i n n g g t t h h e e o o x x y y g g e e n n b b o o t t t t l l e e h h o o l l d d e e r r C C A A U U T T I I O O N N •...
Page 73
F F i i g g u u r r e e 2 2 8 8 – – A A t t t t a a c c h h i i n n g g t t h h e e d d e e f f i i b b r r i i l l l l a a t t o o r r t t r r a a y y O O p p e e r r a a t t i i n n g g t t h h e e d d e e f f i i b b r r i i l l l l a a t t o o r r t t r r a a y y o o p p t t i i o o n n To operate the defibrillator tray: 1.
Page 74
C C l l e e a a n n i i n n g g a a n n d d d d i i s s i i n n f f e e c c t t i i n n g g w w i i t t h h S S i i d d e e K K i i c c k k For United States only. Confirm availability for your configuration or region. Call Stryker Customer Service: 1-800-327- 0770.
Page 75
C C l l e e a a n n i i n n g g C C A A U U T T I I O O N N • Do not clean, disinfect, service, or perform maintenance while the product is in use. •...
Page 76
D D i i s s i i n n f f e e c c t t i i n n g g C C A A U U T T I I O O N N • Do not clean, disinfect, service, or perform maintenance while the product is in use. •...
Page 77
Remove the product from service before you perform the preventive maintenance inspection. Check all items listed during annual preventive maintenance for all Stryker Medical products. You may need to perform preventive maintenance checks more often based on your level of product usage. Service only by qualified personnel.
Page 78
Z Z o o o o m m drives forward and backward when throttle is rotated option Z Z o o o o m m throttle returns to neutral position when released and Z Z o o o o m m motion stops option Product serial number: Completed by: Date:...
Page 79
W W i i r r e e l l e e s s s s n n o o t t i i f f i i c c a a t t i i o o n n s s For product equipped with optional wireless communication technology, these statements apply to the countries as indicated: C C o o u u n n t t r r y y...
Page 80
C C o o u u n n t t r r y y N N o o t t i i f f i i c c a a t t i i o o n n Thailand เครื ่ อ งวิ ท ยุ ค มนาคมนี ้ ม ี อ ั ต ราการดู ด กลื น พลั ง งานจำเพาะ (Specific Absorption Rate - SAR) อั น เนื ่ อ งมาจาก เครื...
Page 81
E E M M C C i i n n f f o o r r m m a a t t i i o o n n W W A A R R N N I I N N G G •...
Page 82
G G u u i i d d a a n n c c e e a a n n d d m m a a n n u u f f a a c c t t u u r r e e r r ’ ’ s s d d e e c c l l a a r r a a t t i i o o n n - - e e l l e e c c t t r r o o m m a a g g n n e e t t i i c c i i m m m m u u n n i i t t y y Main power quality should Electrostatic fast transient/ ±2 kV for power supply lines...
Page 83
Portable and mobile RF communications equipment should follow the guidance in the table titled “Recommended separation distances between portable and mobile RF communication equipment and the P P r r o o C C u u i i t t y y bed series.”...
Page 84
R R e e c c o o m m m m e e n n d d e e d d s s e e p p a a r r a a t t i i o o n n d d i i s s t t a a n n c c e e s s b b e e t t w w e e e e n n p p o o r r t t a a b b l l e e a a n n d d m m o o b b i i l l e e R R F F c c o o m m m m u u n n i i c c a a t t i i o o n n e e q q u u i i p p m m e e n n t t a a n n d d t t h h e e P P r r o o C C u u i i t t y y b b e e d d s s e e r r i i e e s s The P P r r o o C C u u i i t t y y bed series is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled.