P P r r o o C C u u i i t t y y ™ ™ B B e e d d S S e e r r i i e e s s O O p p e e r r a a t t i i o o n n s s M M a a n n u u a a l l 3009 3009-009-001 Rev B.0 2020/08...
Page 3
Fuse rating Non-ionizing radiation Catalogue number Serial number European medical device CE mark 2797 Authorized representative in the European Community For US Patents see www.stryker.com/patents Manufacturer Safe working load Mass of equipment NAWI Class IIII Maximum patient weight 3009-009-001 Rev B.0...
Page 4
Adult patient Alternating current Direct current Duty cycle of product ≤2m ≥18m Unit provides terminal for connection of a potential equalization conductor. The potential equalization conductor provides direct connection between the unit and potential equalization busbar of the electrical installation. Protective earth ground Protection from liquid splash I I P P X X 4 4...
Page 5
T T a a b b l l e e o o f f C C o o n n t t e e n n t t s s Warning/Caution/Note Definition ..........................3 Summary of safety precautions ..........................3 Introduction ................................6 Product description ..............................6 Indications for use..............................6 Clinical benefits ..............................7...
Page 6
Footboard control panel - Menu..........................36 Accessories and parts ............................38 Raising or lowering the IV pole (option)........................38 Attaching or removing the patient helper (option) .....................39 Adjusting the patient helper (option) ........................40 Attaching the oxygen bottle holder..........................41 Cleaning and disinfecting with wipes ........................42 Cleaning ..................................43...
Page 7
This product is equipped with a hospital-grade plug for protection against electric shock hazard. • Always use a Stryker supplied interface cable. Use of any other cable may cause the product to not function as intended, which may result in patient or user injury.
Page 8
• Do not apply the brakes to slow or stop the product while the product is in motion. • Always unplug the power cord before you transport the product. • Always release the brakes before you transport the product. Do not transport the product with the brakes applied. •...
Page 9
• Do not load an individual IV pole hook above the safe working load of 8.8 lb (4 kg). • Always secure the lifting pole in the mounting bracket before you adjust the patient helper. • Always check that the patient helper mounting bracket is secure before use. •...
Page 10
P P r r o o d d u u c c t t d d e e s s c c r r i i p p t t i i o o n n The Stryker Model 3009 P P r r o o C C u u i i t t y y ™ bed series is a powered, adjustable hospital bed used in combination with a patient support surface.
Page 11
Stryker beds that are verified and validated with the i i B B e e d d W W i i r r e e l l e e s s s s software and is not intended to...
Page 12
120 VAC, 60 Hz, 8A Battery voltage 12 VDC, 1.2 Ah (x2) (Stryker part number: 700000341245) N N o o t t e e - - Always replace with Stryker approved batteries. Duty cycle 2 minutes ON, 18 minutes OFF Application environments...
Page 13
O O p p e e r r a a t t i i o o n n S S t t o o r r a a g g e e a a n n d d t t r r a a n n s s p p o o r r t t a a t t i i o o n n E E n n v v i i r r o o n n m m e e n n t t a a l l c c o o n n d d i i t t i i o o n n s s 95 °F 140 °F...
Page 14
SHA-1 and SHA-2 server side certificate recognition for Hash function compatibility PEAP-MSCHAP - v2 2.4 GHz: All channels supported Channel plan 5 GHz: All channels supported (Recommend against the use of DFS and ISM channels) Leverage hospital SSID Support for 802.11r Other Support for Cisco CCX (Fast roaming) S S p p e e c c i i f f i i c c a a t t i i o o n n - - C C h h i i p p s s e e t t Q Q C C A A 9 9 3 3 7 7 7 7 - - 3 3 ( ( Q Q u u a a l l c c o o m m m m A A t t h h e e r r o o s s ) )
Page 15
P P r r o o d d u u c c t t i i l l l l u u s s t t r r a a t t i i o o n n F F i i g g u u r r e e 1 1 – – M M o o d d e e l l 3 3 0 0 0 0 9 9 P P r r o o C C u u i i t t y y b b e e d d s s e e r r i i e e s s Auxiliary outlet Headboard Operator control panel...
Page 16
Competent authority of the European Member State where the user and/or patient is established. To view your operations or maintenance manual online, see https://techweb.stryker.com/. Have the serial number (A) of your Stryker product available when calling Stryker Customer Service or Technical Support. Include the serial number in all written communication.
Page 17
F F i i g g u u r r e e 3 3 – – S S e e r r i i a a l l n n u u m m b b e e r r l l o o c c a a t t i i o o n n D D a a t t e e o o f f m m a a n n u u f f a a c c t t u u r r e e The year of manufacture is the first 2 digits of the serial number.
Page 18
W W A A R R N N I I N N G G • Always use a Stryker supplied interface cable. Use of any other cable may cause the product to not function as intended, which may result in patient or user injury.
Page 19
Installation/Configuration Manual. Contact Stryker Technical Support at (800) 327-0770 with any set up questions. N N o o t t e e - - You must load the wireless connection settings before the device will communicate with the i i B B e e d d Server application.
Page 20
O O p p e e r r a a t t i i o o n n P P l l u u g g g g i i n n g g o o r r u u n n p p l l u u g g g g i i n n g g t t h h e e p p r r o o d d u u c c t t W W A A R R N N I I N N G G •...
Page 21
To charge the battery, connect the product to a hospital grade protective earthed outlet. The battery has a full charge within eight hours. S S t t o o r r i i n n g g t t h h e e p p o o w w e e r r c c o o r r d d W W A A R R N N I I N N G G •...
Page 22
A A p p p p l l y y i i n n g g o o r r r r e e l l e e a a s s i i n n g g t t h h e e b b r r a a k k e e s s W W A A R R N N I I N N G G •...
Page 23
To release S S t t e e e e r r - - L L o o c c k k , depress the red side of the pedal until the pedal is in the neutral position (Figure 9). N N o o t t e e - - To move in any direction, release the S S t t e e e e r r - - L L o o c c k k pedal.
Page 24
F F i i g g u u r r e e 1 1 0 0 – – A A c c t t i i v v a a t t i i n n g g t t h h e e C C P P R R r r e e l l e e a a s s e e h h a a n n d d l l e e R R e e m m o o v v i i n n g g o o r r r r e e p p l l a a c c i i n n g g t t h h e e h h e e a a d d b b o o a a r r d d W W A A R R N N I I N N G G - - Do not use the headboard for CPR support.
Page 25
• To raise the siderail to the highest position, press and hold the siderail release (N) ( Product illustration (page 11)) and rotate the siderail up from either the lowest position or intermediate position (Figure 11). F F i i g g u u r r e e 1 1 1 1 – – S S i i d d e e r r a a i i l l h h i i g g h h e e s s t t p p o o s s i i t t i i o o n n •...
Page 26
When you lower the siderails, listen for a click to indicate that the siderail locks into position. Pull on the siderail to make sure the siderail locks. To lower the siderail to the intermediate position, press the siderail release (N) ( Product illustration (page 11)) and rotate •...
Page 27
F F i i g g u u r r e e 1 1 4 4 – – R R e e s s t t r r a a i i n n t t s s t t r r a a p p t t i i e e - - i i n n l l o o c c a a t t i i o o n n s s S S e e c c u u r r i i n n g g a a F F o o l l e e y y b b a a g g t t o o t t h h e e F F o o l l e e y y b b a a g g h h o o o o k k There are two Foley bag hooks under the foot section (A) (Figure 15), one on either side of the product.
Page 28
N N o o t t e e - - Nurse call requires a connection between the product and an applicable input (patient station, head wall, or docker station). See Setting up wired nurse call communication (page 14). C C o o n n n n e e c c t t i i n n g g p p e e r r i i p p h h e e r r a a l l e e q q u u i i p p m m e e n n t t t t o o t t h h e e a a u u x x i i l l i i a a r r y y o o u u t t l l e e t t W W A A R R N N I I N N G G •...
Page 29
Bed exit Press and hold to cancel bed exit alarm Nurse call Activates nurse call Gatch up Raises the Gatch Gatch down Lowers the Gatch Fowler up Raises the Fowler Fowler down Lowers the Fowler Fowler 30°+ Illuminates when the Fowler is 30°+ Bed height up Raises the litter Bed height down...
Page 30
Places the product into a position for patient ingress or egress Patient stand assist N N o o t t e e - - The patient ingress or egress button disables bed exit. Cardiac chair position Places the product into the cardiac chair position Vascular position Places the product into the vascular position Places the product into the Trendelenburg position...
Page 31
Nurse call Activates nurse call Fowler up Raises the Fowler Fowler down Lowers the Fowler Gatch up Raises the Gatch Gatch down Lowers the Gatch P P e e n n d d a a n n t t , , b b a a s s i i c c ( ( o o p p t t i i o o n n ) ) W W A A R R N N I I N N G G •...
Page 32
P P e e n n d d a a n n t t , , a a d d v v a a n n c c e e d d ( ( o o p p t t i i o o n n ) ) W W A A R R N N I I N N G G •...
Page 33
Mute Mutes the TV/radio TV/radio Turns the TV/radio on or off Closed captions Turns closed captions on or off Room light Turns the room light on or off Reading light Turns the reading light on or off F F o o o o t t b b o o a a r r d d c c o o n n t t r r o o l l p p a a n n e e l l - - H H o o m m e e The home functions are located on the touch panel of the footboard.
Page 34
F F o o o o t t b b o o a a r r d d c c o o n n t t r r o o l l p p a a n n e e l l - - P P o o s s i i t t i i o o n n W W A A R R N N I I N N G G •...
Page 35
F F o o o o t t b b o o a a r r d d c c o o n n t t r r o o l l p p a a n n e e l l - - M M o o t t i i o o n n l l o o c c k k W W A A R R N N I I N N G G •...
Page 36
Returns to the H H o o m m e e screen ( Footboard control panel - Home Home (page 29)) Displays the S S c c a a l l e e I I n n f f o o screen Info Select to save the current displayed weight ( Weighing a Save weight...
Page 37
W W e e i i g g h h i i n n g g a a p p a a t t i i e e n n t t W W A A R R N N I I N N G G - - Do not use the scale system reading as a reference for medical treatment. The scale system assists only to monitor the patient’s weight variation.
Page 38
2. On the S S c c a a l l e e screen, press the A A d d d d E E q q u u i i p p . . button (E) ( Footboard control panel - Scale (page 31)). 3.
Page 39
3. Press the B B e e d d E E x x i i t t button (B) on the footboard control panel to arm bed exit ( Footboard control panel - Home (page 29)). 4. Select the desired zone (B, C, D) ( Footboard control panel - Bed exit (page 34)). If you change the parameter conditions for bed exit: •...
Page 40
E E n n a a b b l l i i n n g g o o r r d d i i s s a a b b l l i i n n g g i i B B e e d d W W a a t t c c h h To enable i i B B e e d d W W a a t t c c h h : 1.
Page 41
Settings Displays settings Displays advanced settings Advanced N N o o t t e e - - Press and hold the S S e e t t t t i i n n g g s s button (F) to display the A A d d v v a a n n c c e e d d button.
Page 42
A A c c c c e e s s s s o o r r i i e e s s a a n n d d p p a a r r t t s s These accessories and parts may be available for use with your product. Confirm availability for your configuration or region.
Page 43
F F i i g g u u r r e e 1 1 6 6 – – T T w w o o - - s s t t a a g g e e I I V V p p o o l l e e s s h h o o w w n n A A t t t t a a c c h h i i n n g g o o r r r r e e m m o o v v i i n n g g t t h h e e p p a a t t i i e e n n t t h h e e l l p p e e r r ( ( o o p p t t i i o o n n ) ) W W A A R R N N I I N N G G - - Always use two people to attach or remove the patient helper.
Page 44
F F i i g g u u r r e e 1 1 7 7 – – A A t t t t a a c c h h i i n n g g o o r r r r e e m m o o v v i i n n g g t t h h e e p p a a t t i i e e n n t t h h e e l l p p e e r r Reverse steps to remove the patient helper.
Page 45
F F i i g g u u r r e e 1 1 8 8 – – A A d d j j u u s s t t i i n n g g t t h h e e p p a a t t i i e e n n t t h h e e l l p p e e r r A A t t t t a a c c h h i i n n g g t t h h e e o o x x y y g g e e n n b b o o t t t t l l e e h h o o l l d d e e r r C C A A U U T T I I O O N N •...
Page 46
For United States only. Confirm availability for your configuration or region. Call Stryker Customer Service: 1-800-327- 0770. Stryker’s preferred wipes (2060-000-001 6'' x 10'' or 2060-000-002 9'' x 12'') include the following active ingredients: • n-Alkyl (60% C14, 30% C16, 5% C12, 5% C18) dimethyl benzyl ammonium chloride - 0.154% •...
Page 47
C C l l e e a a n n i i n n g g C C A A U U T T I I O O N N • Do not clean, disinfect, service, or perform maintenance while the product is in use. •...
Page 48
D D i i s s i i n n f f e e c c t t i i n n g g C C A A U U T T I I O O N N • Do not clean, disinfect, service, or perform maintenance while the product is in use. •...
Page 49
Remove the product from service before you perform the preventive maintenance inspection. Check all items listed during annual preventive maintenance for all Stryker Medical products. You may need to perform preventive maintenance checks more often based on your level of product usage. Service only by qualified personnel.
Page 51
W W i i r r e e l l e e s s s s n n o o t t i i f f i i c c a a t t i i o o n n s s For product equipped with optional wireless communication technology, these statements apply to the countries as indicated: C C o o u u n n t t r r y y...
Page 52
E E M M C C i i n n f f o o r r m m a a t t i i o o n n W W A A R R N N I I N N G G •...
Page 53
G G u u i i d d a a n n c c e e a a n n d d m m a a n n u u f f a a c c t t u u r r e e r r ’ ’ s s d d e e c c l l a a r r a a t t i i o o n n - - e e l l e e c c t t r r o o m m a a g g n n e e t t i i c c i i m m m m u u n n i i t t y y Main power quality should Electrostatic fast transient/ ±2 kV for power supply lines...
Page 54
Portable and mobile RF communications equipment should follow the guidance in the table titled “Recommended separation distances between portable and mobile RF communication equipment and the 3009 P P r r o o C C u u i i t t y y bed series.”...
Page 55
R R e e c c o o m m m m e e n n d d e e d d s s e e p p a a r r a a t t i i o o n n d d i i s s t t a a n n c c e e s s b b e e t t w w e e e e n n p p o o r r t t a a b b l l e e a a n n d d m m o o b b i i l l e e R R F F c c o o m m m m u u n n i i c c a a t t i i o o n n e e q q u u i i p p m m e e n n t t a a n n d d t t h h e e 3 3 0 0 0 0 9 9 P P r r o o C C u u i i t t y y b b e e d d s s e e r r i i e e s s The 3009 P P r r o o C C u u i i t t y y bed series is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled.
Page 56
Stryker Medical 3800 E. Centre Avenue Portage, MI 49002 3009-009-001 Rev B.0 2020/08 WCR: AA.21...