Now that the site for your Protector Work Station is properly prepared, you
are ready to unpack, inspect, install, and validate your system. Read this
chapter to learn how to:
 Unpack and move the enclosure.
 Set up the enclosure with the proper supporting structure and work
surface.
 Connect to an exhaust system (only applicable for remote blower).
 Installation of carbon filters (only applicable for integral blower
exhausted to the room).
 Connect the electrical supply.
 Set the face velocity with the speed control adjustment, only
applicable for integral blower.
 Arrange validation for the enclosure.
 Seal the enclosure to the work surface.
Depending upon which model you are installing, you may need common
mechanical and electrical installation tools in addition to wrenches,
ratchets, sockets, a nut driver set, a flat-blade screwdriver, a Phillips
screwdriver, and a carpenter level to complete the instructions in this
chapter.
Each enclosure weighs either 85 or 105 lbs. each (37 or
!
46 kg). The shipping container allows for lifting with a
mechanical lift truck or floor jack. If you must lift the
enclosure manually, follow safe-lifting guidelines.
not lift by the front air foil