Download Print this page

Endress+Hauser Proline Promass E 100 Operating Instructions Manual page 95

Coriolis flowmeter ethernet/ip
Hide thumbs Also See for Proline Promass E 100:

Advertisement

Proline Promass E 100 EtherNet/IP
Diagnostic information
No.
274 Main electronic failure
Status signal
Diagnostic behavior
Diagnostic information
No.
283 Memory content
Status signal
Diagnostic behavior
Diagnostic information
No.
311 Electronic failure
Status signal
Diagnostic behavior
Diagnostic information
No.
311 Electronic failure
Status signal
Diagnostic behavior
Diagnostic information
No.
382 Data storage
Status signal
Diagnostic behavior
Diagnostic information
No.
383 Memory content
Status signal
Diagnostic behavior
Endress+Hauser
Short text
S
Warning
Short text
F
Alarm
Short text
F
Alarm
Short text
M
Warning
Short text
F
Alarm
Short text
F
Alarm
Diagnostics and troubleshooting
Remedy instructions
Change electronic
Remedy instructions
1. Reset device
2. Contact service
Remedy instructions
1. Reset device
2. Contact service
Remedy instructions
1. Do not reset device
2. Contact service
Remedy instructions
1. Insert DAT module
2. Change DAT module
Remedy instructions
1. Restart device
2. Check or change DAT module 3.
Contact service
Coding of diagnostic
information (hex)
• 0x80000CE
• 0x80000CF
Coding of diagnostic
information (hex)
0x100016F
Coding of diagnostic
information (hex)
0x10000E1
Coding of diagnostic
information (hex)
0x40000E2
Coding of diagnostic
information (hex)
0x100016D
Coding of diagnostic
information (hex)
0x100016E
95