Instructions for use MULTIflex LED coupling 465 LED - 1.007.3197
7 Reprocessing steps in accordance with ISO 17664 | 7.4 Automated reprocessing
For effective reprocessing, automated internal and external cleaning as well as
automated internal and external disinfection with a cleaning and disinfection
unit in accordance with EN ISO 15883-1 is required.
7.3.2 Manual internal and external disinfection
This product is not designed for manual internal and external disinfection.
For effective reprocessing, automated internal and external cleaning as well as
automated internal and external disinfection with a cleaning and disinfection
unit in accordance with EN ISO 15883-1 is required.
For effective reprocessing, the inside of the device must be subjected to auto-
mated cleaning in a cleaning and disinfection unit in accordance with EN ISO
15883-1.
4 Manual external disinfection is permissible only as an occupational safety
measure (personal protection measure).
Never reprocess the medical device with chloride-containing products.
Malfunction and material damage.
4 Reprocess it in a washer disinfector only.
KaVo recommends the following products based on the compatibility of the ma-
terials. The microbiological efficacy must be ensured by the disinfectant manu-
facturer and proven by an expert opinion.
Approved disinfectants:
▪ KaVo Wipes, CaviWipes and CaviCide made by Metrex
▪ Mikrozid AF made by Schülke & Mayr (Liquid or wipes)
▪ FD 322 made by Dürr
Requisite consumables:
▪ Cloths for wiping the medical device.
4 Spray the disinfectant on a cloth, then wipe down the medical device and al-
low the disinfectant to act according to the instructions of the disinfectant
manufacturer.
4 Follow the instructions for use of the disinfectant.
7.3.3 Manual drying
This product is not designed for manual drying.
For effective reprocessing, automated internal and external cleaning as well as
automated internal and external disinfection with a cleaning and disinfection
unit in accordance with EN ISO 15883-1 is required.
7.4 Automated reprocessing
Incomplete disinfection.
Infection hazard.
4 Only use disinfection procedures that are verified to be bactericidal, fungi-
cidal and virucidal.
CAUTION
WARNING
19 / 28